Perceptually optimized NPS-based deep learning framework for PET image denoising.
Erick Almeida de Souza, Bruno Henrique Lira Dos Anjos, Vinicius de Oliveira Menezes, Cleiton Cavalcante Queiroz, Felipe Alves Mourato, Maria Eduarda Duarte De Mello Flamini, Aline Lopes Garcia Leal, Tsang Ing Ren,
Klar Inteligência Artificial, Recife, PE, Brazil. · Klar Inteligência Artificial, Recife, PE, Brazil; Hospital das Clínicas da Universidade Federal de Pernambuco (HC/UFPE-HU Brasil), Recife, PE, Brazil.
Key points
From the paper's conclusion- Scanner-specific NPS simulation provides a feasible strategy for generating PET denoising training data when LM acquisitions are limited. Larger independent and external studies are required to establish generalizability and clinical applicability.
Abstract
PURPOSE: To develop and evaluate a scanner-specific noise power spectrum (NPS)-based framework for deep learning PET image denoising under reduced-count conditions. METHODS: Retrospective whole-body 18F-FDG PET/CT datasets from 250 subjects and three scanners were used to train a perceptually optimized 2.5D U-Net. Data were partitioned at the patient level into scanner-specific training, validation, and independent testing subsets. Training strategies included NPS-simulated data (DL-NPS), list-mode reconstructions (DL-LM), and their combination (DL-all). Image quality was evaluated using SSIM, PSNR, RMSE, liver coefficient of variation (CV), lesion contrast-to-noise ratio (CNR), and SUV agreement. CNR-based lesion visibility was assessed using the Rose Criterion, while three nuclear medicine physicians independently performed blinded image-quality assessments. RESULTS: DL-NPS achieved competitive performance relative to LM-trained models across scanner configurations. Denoising improved lesion CNR by 15-38% at 25% and 50% counts, increasing the proportion of lesions meeting the Rose visibility threshold by up to 13.5% after denoising. Liver CV decreased toward standard-count reference levels. Bland-Altman analysis showed limited SUV bias (absolute mean bias ≤0.17, 95% LoA within ±1.1). Reader assessments generally favored DL-denoised images, with significant paired improvements for Biograph Horizon but not Discovery 710 after multiplicity correction. CONCLUSION: Scanner-specific NPS simulation provides a feasible strategy for generating PET denoising training data when LM acquisitions are limited. Larger independent and external studies are required to establish generalizability and clinical applicability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Acquisition time/dose reduction in pediatric PET imaging using patch-based deep learningEJNMMI Physics · 2026
- Meta-PET: A graininess-modulated test-time adaptive network for low-count PET denoising.Medical image analysis · 2026
- Deep progressive reconstruction for low-dose ¹⁸F-FDG PET/CT imaging in pediatric patients.Annals of nuclear medicine · 2026
- Deep learning denoising and positron range correction of clinical [68Ga]Ga-DOTA-TOC PET images.Zeitschrift fur medizinische Physik · 2026
- Deep learning-based approaches for attenuation correction in [18F]FDG PET: Current advances and future directions.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Time-aware frame-aligned diffusion-inspired model for late dynamic PET frame prediction from early [11C] raclopride.Physics in medicine and biology · 2026