Severe Hyperglycemia due to Calibration Misuse in an Eversense-365 Implantable Continuous Glucose Monitoring System.
Jennifer Moeller
Division of Diabetes, Endocrinology and Metabolism, Department of Internal Medicine, University of Nebraska Medical Center, Nebraska Medicine, Omaha, Nebraska.
Key points
From the paper's conclusion- Improper calibration of implantable CGM systems can lead to clinically significant inaccuracies and delayed recognition of hyperglycemia. Reinforcing patient education, adhering to recommended calibration protocols, and selecting patients thoughtfully are essential to ensure the safe and effective use of these devices.
Abstract
BACKGROUND: Continuous glucose monitoring (CGM) systems have transformed diabetes management by improving glycemic outcomes and providing real-time glucose data. Accuracy is commonly assessed by using metrics such as mean absolute relative difference. Implantable CGM systems, including Eversense-365, have demonstrated high accuracy and safety when used in accordance with recommended protocols. OBJECTIVE: To describe a case of clinically significant CGM inaccuracy resulting from improper user calibration and to highlight associated safety implications. CASE REPORT: An adult with a 24-year history of type 1 diabetes complicated by microvascular and macrovascular disease, hypoglycemia unawareness, and mild cognitive impairment was using an implantable CGM system. Initially, sensor readings closely matched capillary glucose values when appropriate fingerstick calibration techniques were followed. However, the patient subsequently used CGM-derived glucose values in place of capillary measurements for calibration. This practice led to progressively inaccurate sensor readings, yielding falsely reassuring glucose values and delaying recognition of clinically significant hyperglycemia. DISCUSSION: This case highlights a critical safety consideration in implantable CGM systems that rely on user-dependent calibration. Even with otherwise high-performing technology, improper calibration practices can significantly compromise accuracy. As CGM systems become increasingly integrated into automated insulin delivery platforms, user-related factors may introduce risk if proper technique is not consistently followed. CONCLUSION: Improper calibration of implantable CGM systems can lead to clinically significant inaccuracies and delayed recognition of hyperglycemia. Reinforcing patient education, adhering to recommended calibration protocols, and selecting patients thoughtfully are essential to ensure the safe and effective use of these devices.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Performance of the i-SENS CareSens Air CGM System in Compliance with the European CGM (eCGM) Clinical Performance Criteria: A Pooled Analysis.Diabetes therapy : research, treatment and education of diabetes and related disorders · 2026
- Real-World Data from U.S. Commercial Users of the 365-Day Implantable Eversense Continuous Glucose Monitoring System, Including with the Twiist Automated Insulin Delivery System Integration.Diabetes technology & therapeutics · 2026
- Determinants of Continuous Glucose Monitoring Accuracy in Critically Ill Patients Requiring Mechanical Ventilation and Vasopressor Support.Diabetes technology & therapeutics · 2026
- Continuous glucose monitoring in peritoneal dialysis: What the evidence supports and where it stops.Peritoneal dialysis international : journal of the International Society for Peritoneal Dialysis · 2026
- International clinical opinion on transparency, standardisation, and calibration alignment in the performance evaluation of systems for continuous glucose monitoringDiabetes Obesity and Metabolism · 2026
- Toward standardized continuous glucose monitoring validation: lessons from the Sinocare iCan i3 accuracy study.Frontiers in clinical diabetes and healthcare · 2026