Validation of Automated 3D MRI Knee Cartilage Measurements.
Mirela Kostadinova, Iman Harrou, Scherwin Mahmoudi, Simon Bernatz, Elena Hoehne, Philipp Reschke, Aynur Goekduman, Leona S Alizadeh,
Clinic for Radiology and Nuclear Medicine, University Hospital Frankfurt, Goethe University, Frankfurt am Main, Germany (M.K., I.H., S.M. S.B., E.H., P.R., A.G., L.S.A., L.D.G., C.B., K.E., T.J.V., and I.Y. ). Electronic address: kostadinova.a.mirela@gmail.com. · Clinic for Radiology and Nuclear Medicine, University Hospital Frankfurt, Goethe University, Frankfurt am Main, Germany (M.K., I.H., S.M. S.B., E.H., P.R., A.G., L.S.A., L.D.G., C.B., K.E., T.J.V., and I.Y. ).
Key points
From the paper's conclusion- Following a simple global calibration, automated cartilage thickness measurements achieve overall acceptable agreement with manual reference measurements that is comparable to inherent intersequence variability. This supports the clinical acceptability of the automated approach as an efficient and reproducible alternative to manual segmentation.
Abstract
RATIONALE AND OBJECTIVES: To assess the agreement between automated cartilage thickness measurements derived from three-dimensional (3D) dual-echo steady-state (DESS) magnetic resonance imaging and manual radiologist measurements. MATERIALS AND METHODS: In 111 patients, cartilage thickness measurements across 21 knee subregions were compared between an automated segmentation software and manual radiologist measurements on 3D DESS and proton density (PD) fat-suppressed sequences. Agreement was assessed using Bland-Altman analysis, Pearson correlation, and intraclass correlation coefficient (ICC). A post-hoc calibration approach was applied to both comparisons. Intersequence variability between PD manual and DESS manual measurements served as a benchmark. RESULTS: The automated method showed a systematic overestimation relative to the PD manual (bias +0.7 mm) and the DESS manual (bias +0.9 mm). After calibration, bias was eliminated. Postcalibration ICC improved substantially (PD: from 0.29-0.61; DESS: from 0.18-0.67), reaching levels comparable to the intersequence benchmark (ICC = 0.73). Per-region analysis revealed the greatest variability in patellar and lateral tibial subregions. CONCLUSION: Following a simple global calibration, automated cartilage thickness measurements achieve overall acceptable agreement with manual reference measurements that is comparable to inherent intersequence variability. This supports the clinical acceptability of the automated approach as an efficient and reproducible alternative to manual segmentation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- AUTOMATIC HIGH-RESOLUTION CARTILAGE THICKNESS MAPPING AND SCORING: EXTENSION TO STANDARD CLINICAL MRI - DATA FROM THE LAUSANNE KNEE STUDY (LKS).Osteoarthritis imaging · 2026
- Clinical Validation of Artificial Intelligence (AI)-based Cartilage Segmentation Predicting Knee Replacement.Arthritis care & research · 2026
- Development, Evaluation, and Multicenter Clinical-Trial Application of an Artificial Intelligence-Assisted MRI Method for Quantitative Knee Cartilage MorphometryarXiv · 2026
- Utility of ultrasound as a diagnostic tool for measuring knee cartilage thickness in patients with osteoarthritis: correlation with MRIThe Egyptian Journal of Radiology and Nuclear Medicine · 2026
- Artificial intelligence-based segmentation and quantitative assessment of knee cartilage in osteoarthritis: a systematic review and meta-analysis.International orthopaedics · 2026
- 3D QUANTITATIVE MEASUREMENTS OF JOINT SPACING BASED ON SEMI-AUTOMATIC AND AUTOMATIC SEGMENTATION, CONDUCTED 36 MONTHS APART, USING COMPUTED TOMOGRAPHY IMAGES.Osteoarthritis imaging · 2026