Anaesthesia, critical care & pain medicine

I-gelTM Plus implements higher oropharyngeal leak pressure in the lateral decubitus position than original i-gelTM: a prospective randomized controlled trial.

Shingo Kawashima, Masanori Haba, Atsushi Kobayashi, Tetsuro Kimura, Hiroyuki Kinoshita

Department of Anesthesiology, Seirei Mikatahara General Hospital, Hamamatsu, Japan; Departments of Anesthesiology and Intensive Care, Hamamatsu University School of Medicine, Hamamatsu, Japan. · Department of Anesthesiology, Hidaka Hospital, Gobo, Japan.

Randomised · n=42

Key points

From the paper's conclusion
  • The i-gelTM Plus implements higher oropharyngeal leak pressure in the lateral decubitus position than the original i-gelTM. Our results indicate that the i-gelTM Plus may be a better choice as a supraglottic airway device in the lateral decubitus position than the original i-gelTM in paralyzed patients.

Abstract

BACKGROUND: The supraglottic airway device i-gelTM Plus (Intersurgical Ltd., Berkshire, UK) may provide more secure ventilation than the original i-gelTM (Intersurgical Ltd., Berkshire, UK). This prospective randomized trial was conducted to evaluate the performance of i-gelTM Plus compared with the original i-gelTM in patients in the lateral decubitus position under general anesthesia with a neuromuscular blocking drug. METHODS: We inserted two types of i-gelTM in the supine position in adult patients undergoing general anesthesia with 0.6 mg/kg IV rocuronium. The primary outcome was oropharyngeal leak pressure in the lateral position. Secondary outcomes included expiratory tidal volume and glottic view in that position. RESULTS: We analyzed 84 surgical patients in the i-gelTM Plus [n = 42] and original i-gelTM [n = 42] groups. The oropharyngeal leak pressure and expiratory tidal volume at the oropharyngeal leak pressure in the supine position were similar between the two groups. The lateral decubitus position demonstrated higher oropharyngeal leak pressure (P < 0.001) and expiratory tidal volume at oropharyngeal leak pressure (P < 0.001) in the i-gelTM Plus group (median [interquartile range], 22 [20-25] cmH2O and 694 [602-750] ml, respectively) than in the original i-gelTM group (15 [14-18] cmH2O and 470 [420-634] ml, respectively). However, fiberoptic glottis view grade scores did not differ between the two groups. CONCLUSION: The i-gelTM Plus implements higher oropharyngeal leak pressure in the lateral decubitus position than the original i-gelTM. Our results indicate that the i-gelTM Plus may be a better choice as a supraglottic airway device in the lateral decubitus position than the original i-gelTM in paralyzed patients. REGISTRATION: University Hospital Medical Information Network (UMIN000061937).

Read the paper