Surface plasmon resonance imaging detection of microRNA using gold-magnetite nanoreporters with optimized probe interfaces.

Jihao Cheng, Chaoyi Zeng, Shangzhong Jin

College of Optical and Electronic Technology, China Jiliang University, Hangzhou 310018, China. · College of Optical and Electronic Technology, China Jiliang University, Hangzhou 310018, China. Electronic address: Jinsz@cjlu.edu.cn.

Key points

From the paper's conclusion
  • These results demonstrate improved SPRi responses through capture and reporter interface optimization and the use of composite nanoreporters.

Abstract

Direct SPRi detection of short-chain miRNAs remains challenging because their hybridization can produce a relatively small interfacial optical response. We developed a sandwich SPRi assay using gold nanoparticle/magnetite (AuNPs/Fe3O4) composite nanoreporters, with miRNA-21 as the model target. The capture and reporter interfaces were optimized by examining probe immobilization conditions, capture-probe concentration, gold nanoparticle size, and polyT spacer length. Freeze-assisted immobilization shortened the immobilization step from 24 to 2 h while maintaining comparable SPRi responses. A capture-probe concentration of 50 μM gave the highest response among the concentrations tested. The 50 nm gold nanoparticles produced higher responses than the 20 nm particles, and a reporter probe containing a 10-thymidine spacer performed best among the tested spacer lengths. Composite nanoreporters further increased the response relative to gold nanoparticle reporters alone at concentrations tested in both assays. The resulting assay showed a log-linear relationship between SPRi response and miRNA-21 concentration from 5 fM to 250 pM (R2 = 0.94769). A shared blank-derived response threshold gave an estimated mean limit of detection (LOD) of 3.27 ± 0.19 fM based on three independent experiments (n = 3), obtained by extrapolation below the lowest calibration concentration (5 fM). At 100 fM, miRNA-21 produced a higher response than the three tested non-target sequences. These results demonstrate improved SPRi responses through capture and reporter interface optimization and the use of composite nanoreporters.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Surface plasmon resonance imaging detection of microRNA using gold-magnetite nanoreporters with optimized probe interfaces — Science Skim