Measurement of Macular Pigment in Singaporeans: Macular Metrics Densitometer Versus Macular Pigment Screener.
Kumari Neelam, Joanna H Cher, Chalani U Ubeynarayana, Ivan Y-F Leung, Ian J Murray
Department of Ophthalmology and Visual Sciences, Khoo Teck Puat Hospital, Singapore, Singapore. Kumari.neelam@nhghealth.com.sg. · Clinical Research Unit, Khoo Teck Puat Hospital, Singapore, Singapore.
Key points
From the paper's conclusion- MPS II measures MPOD in a reliable and reproducible manner when compared with MMD. However, MPS II consistently recorded lower values than MMD and has a longer recording time, particularly in those with higher MPOD. Moreover, MPS II failed to reach necessary level of agreement with MMD to allow interchangeability on an individual basis.
Abstract
PURPOSE: To compare macular pigment optical density (MPOD) obtained using a macular pigment screener (MPS II) and macular metrics densitometer (MMD). METHODS: Following details were collected in 65 healthy Singaporeans: sociodemographic, lifestyle information, height and weight, MPOD using MMD and MPS II including time taken to obtain measurements, and blood sample for estimation of serum lutein (L) and zeaxanthin (Z). MPOD agreement between MMD and MPS II was assessed using Bland-Altman plots and 95% limits of agreement. Intraclass correlation coefficient (ICC) was calculated to assess absolute concordance between the two instruments. RESULTS: Mean (±SD) age of the population was 27.92 years (±8.60) with slight preponderance of females. Mean MPOD measured with MPS II (0.48 ± 0.17) was significantly lower than MMD (0.57 ± 0.21). MPOD values recorded using MMD and MPS II were correlated (r = 0.72, p < 0.0001). Using Bland-Altman plots, upper and lower 95% limits of agreement were 0.370 and -0.194, respectively. Concordance of MPOD measured by MMD and MPS II using ICC was 0.64 (95% CI: 0.343-0.798). MPOD recording time was significantly longer using MPS II than MMD (p < 0.05). Recording time increased significantly in subjects with higher MPOD using MPS II (p = 0.004), whereas MPOD recording time did not differ with MMD (p = 0.99). A weak correlation was observed between serum L and MPOD measured with MPS II (r = 0.31, p = 0.01) and MMD (r = 0.24, p = 0.06). Coefficients of repeatability for MMD and MPS II were 0.21 and 0.06, respectively. CONCLUSIONS: MPS II measures MPOD in a reliable and reproducible manner when compared with MMD. However, MPS II consistently recorded lower values than MMD and has a longer recording time, particularly in those with higher MPOD. Moreover, MPS II failed to reach necessary level of agreement with MMD to allow interchangeability on an individual basis.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Macular Pigment Optical Density Is Not Associated with Contrast Sensitivity When Measured with 520 nm Monochromatic Stimuli.Vision (Basel, Switzerland) · 2026
- The functional significance of human macular pigment: Optical, visual and neural effects.Progress in retinal and eye research · 2026
- Macular Pigment and Retinal Reflectance in a Feasibility Study of Multiple Sclerosis: Toward Novel Biomarkers of Hidden Disease Processes.Investigative ophthalmology & visual science · 2026
- Agreement, Reliability, and Feasibility of Melbourne Rapid Fields Tablet Perimetry Compared with Humphrey Field Analyzer in Glaucoma.Diagnostics (Basel, Switzerland) · 2026
- Macular Pigment Changes in Patients with Diabetes Mellitus and Without Diabetic Retinopathy.Ophthalmology and therapy · 2026
- A comparative analysis of pupillometry measurements obtained from different devices featuring static and dynamic capabilities.Documenta ophthalmologica. Advances in ophthalmology · 2026