An Experience of Orthopedic Implant-Related Adverse Events at a Regional Training Center in Central India.
Ahmad Najmi, Shilpa N Kaore, Rehan-Ul-Haq, Manish Dwivedi, S Balakrishnan, Mohammad Faizan Khan, Chirag Agrawal
Department of Pharmacology, All India Institute of Medical Sciences, Saket Nagar, Bhopal, Madhya Pradesh 462020 India. · Department of Orthopaedics, All India Institute of Medical Sciences, Saket Nagar, Bhopal, Madhya Pradesh 462020 India.
Key points
From the paper's conclusion- This study highlights infections as the predominant OIAE, requiring revision surgeries in most cases. A robust materiovigilance system and increased awareness among healthcare providers are essential for timely detection and reporting of OIAEs to improve patient safety.
Abstract
BACKGROUND: Orthopedic implant-associated adverse events (OIAEs) pose a significant challenge to patient safety. Infections, implant failures, allergic reactions, and the need for revision surgeries are among the most frequently reported complications. This study aimed to evaluate the nature and pattern of medical device adverse events (MDAEs) related to orthopedic implants at a Regional Training Center under the Materiovigilance Program of India. METHODS: A retrospective, observational study was conducted using spontaneously reported MDAEs at the Medical Device Monitoring Center (MDMC), at a tertiary care center in Central India, between July 2019 and September 2022. Reports were submitted by healthcare professionals, patients, or their caregivers. Data were collected using the Indian Pharmacopeia Commission (IPC) recommended MDAE reporting form and analyzed using Microsoft Excel and SPSS v1.0.0.1406. Causality was assessed using WHO-UMC criteria. Descriptive statistics (frequency, percentages) were used for analysis. RESULTS: A total of 92 MDAEs were reported, of which 88% were serious. Infections were the leading cause (73%), followed by dislocation (11%), allergic reactions (7%), implant loosening (5%), and metallosis or bent nails (2% each). Revision surgery was required in 76% of cases. The most frequently implicated devices included interlocking nails and tibia locking plates. Causality assessment categorized 58% as probable, 25% as possible, and the remainder as unlikely, unassessable, or not related. CONCLUSION: This study highlights infections as the predominant OIAE, requiring revision surgeries in most cases. A robust materiovigilance system and increased awareness among healthcare providers are essential for timely detection and reporting of OIAEs to improve patient safety.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Phenotypic Characterisation of Resistant Profiles of the Isolates in Orthopaedic Implant Associated Infections: A Cross-sectional StudyJOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH · 2026
- Patient safety in dentistry: Surveillance of adverse events associated with medical devices.Journal of dentistry · 2026
- Microbiological profile and antimicrobial susceptibility pattern in orthopaedic implant associated infections.Iranian journal of microbiology · 2026
- Development and validation of a standardized causality assessment tool for adverse events associated with medical devices.Perspectives in clinical research · 2026
- Analysis of causative factors and preventive strategies for traumatic orthopedic medical malpractice in Shanghai: a high-responsibility perspective.Frontiers in surgery · 2026
- Epidemiology and healthcare burden of implant-associated complications in Germany: a nationwide population-based analysis.European journal of trauma and emergency surgery : official publication of the European Trauma Society · 2026