A Translational Paradigm for Univentricular Heart Disease: Single-Stage Neonatal Repair with Adaptive Control of Cavopulmonary Flow.
Paulo Cesar Duarte Junior, Martin Poulsen Kessler, Alessandro Cavalcanti Lianza, Shirley Ferraz Crispilho, Rudolf Huebner, Hemerson Donizete Pinheiro, Aron José Pazin de Andrade
Department of Bioengineering, Dante Pazzanese Institute of Cardiology, São Paulo, Brazil. paulo.duarte@dantepazzanese.org.br. · ESSS Institute, Florianópolis, Santa Catarina, Brazil.
Key points
From the paper's conclusion- Within the assumptions of this proof-of-concept computational model, the proposed strategy demonstrated hemodynamic feasibility for progressive cavopulmonary flow modulation. By integrating tissue engineering, adaptive flow control, and computational modeling, the proposed concept provides a foundation for future patient-specific, experimental, and translational investigations in the treatment of univentricular heart defects.
Abstract
BACKGROUND: Univentricular heart defects remain one of the greatest challenges in pediatric cardiovascular surgery. Although well-established, the traditional approach consisting of three serial palliative procedures (Norwood, Glenn, and Fontan) entails cumulative risks, repeated hospitalizations, high healthcare costs, and significant psychosocial impact. This study proposes a single-stage neonatal surgical strategy integrating a tissue-engineered vascular graft and a differential flow control valve to enable adaptive and gradual modulation of cavopulmonary flow, avoiding abrupt physiological transitions. To evaluate the hemodynamic feasibility of the proposed concept, a three-dimensional computational fluid dynamics model was developed and analyzed under controlled flow conditions. METHODS: A three-dimensional model of the total cavopulmonary connection was developed to evaluate the hemodynamic feasibility of the surgical proposal. The fenestration diameter was reduced from 14 to 0 mm, representing progressive hemodynamic conditions. Computational fluid dynamics simulations were conducted under standardized physiological conditions, with neonatal pulmonary vascular resistance represented by porous zones. Flow distributions, pressure gradients, and hemodynamic equilibrium points were evaluated. RESULTS: Progressive reduction of fenestration diameter from 14 to 0 mm promoted gradual redistribution of venous flow from predominant right atrial diversion to predominant pulmonary perfusion. Systemic venous pressure increased from 4.2 to 10.6 mmHg as fenestration diameter decreased. An intermediate functional range between 5 and 6 mm provided a balanced distribution between pulmonary perfusion and right atrial diversion, consistent with an intermediate cavopulmonary flow condition. CONCLUSION: Within the assumptions of this proof-of-concept computational model, the proposed strategy demonstrated hemodynamic feasibility for progressive cavopulmonary flow modulation. By integrating tissue engineering, adaptive flow control, and computational modeling, the proposed concept provides a foundation for future patient-specific, experimental, and translational investigations in the treatment of univentricular heart defects.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multiscale Modeling of a Geometrically Tunable Shunt for Growing Pediatric Patients With Single Ventricle Physiology.Artificial organs · 2026
- Hemodynamic optimization of a novel single-puncture dual-lumen coaxial catheter for arteriovenous fistula access: A computational fluid dynamics study.Northern clinics of Istanbul · 2026
- A Novel Ex Vivo Model of Unbalanced Common Atrioventricular Valve: Insights into Biomechanics and Hemodynamics During Single Ventricle Palliation.The Journal of thoracic and cardiovascular surgery · 2026
- From Simulation to Suture Lines: Computational Fluid Dynamic Simulation-Guided Baffle Construction for Complex Biventricular RepairarXiv · 2026
- Two slits good, four slits bad? Assessing the hemodynamic impact of modifications to microvascular plugs in pulmonary flow restriction using a bench-top model.Cardiology in the young · 2026
- Clinically Motivated Redesign of the Pennsylvania State University Child Ventricular Assist Device.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026