Validation of convenient correction methods for collimator-dependent differences in the heart-to-mediastinum ratio from cardiac 123I-MIBG imaging.
Yusuke Inoue, Yutaka Abe, Kohei Mitsui, Hiroki Miyatake, Keita Takahashi, Kei Kikuchi, Eiji Kitamura, Sota Oguro,
Department of Radiological Sciences, Faculty of Health Sciences, Komazawa University, Tokyo, 154- 8525, Japan. inoueys34@gmail.com. · Department of Radiology, Kitasato University Hospital, Sagamihara, Japan.
Key points
From the paper's conclusion- Convenient correction methods using 123I point-source imaging were valid for reducing collimator-dependent differences in the H/M ratio in cardiac 123I-MIBG imaging, demonstrating their applicability to a GE system.
Abstract
OBJECTIVE: High-energy photons emitted by 123I may penetrate the septa of a collimator, thereby degrading quantitative accuracy. Collimator choice affects the evaluation of the heart-to-mediastinum (H/M) ratio in cardiac sympathetic imaging with 123I-metaiodobenzylguanidine (123I-MIBG). We investigated the validity of previously reported convenient methods for correcting collimator-dependent differences in the H/M ratio. METHODS: Twenty patients who underwent cardiac 123I-MIBG imaging for evaluation of neurodegenerative diseases on a gamma camera system provided by GE Healthcare were studied. Anterior chest images were acquired successively with three collimators: the low-energy high-resolution and sensitivity (LEHRS), extended low-energy general-purpose (ELEGP), and medium-energy general-purpose (MEGP) collimators. The H/M ratios were estimated using various correction methods: 123I-dual-window (IDW) correction using the high-energy subwindow, empiric correction based on 123I point-source imaging, combined correction using the high-energy subwindow and point-source imaging, and phantom-based method using imaging of a dedicated phantom. Empiric correction and combined correction had been established using patient data mainly obtained with a system provided by Siemens Healthineers. RESULTS: The uncorrected H/M ratios obtained with the LEHRS collimator correlated closely with those obtained with the MEGP collimator but were substantially smaller, indicating severe underestimation with the LEHRS collimator, attributable to septal penetration. The underestimation remained but decreased after any correction. The H/M ratios after correction were largest with combined correction, followed by empiric correction, phantom-based correction, and IDW correction. The absolute errors after correction were smallest for combined correction, followed by empiric correction, phantom-based correction, and IDW correction. The underestimation was substantially smaller with the ELEGP collimator than with the LEHRS collimator, and any type of correction reduced the underestimation. CONCLUSIONS: Convenient correction methods using 123I point-source imaging were valid for reducing collimator-dependent differences in the H/M ratio in cardiac 123I-MIBG imaging, demonstrating their applicability to a GE system.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Basal coronary microvascular resistance is associated with cardiac sympathetic innervation assessed by 123I-MIBG scintigraphy in patients with heart failureEuropean Heart Journal · 2026
- Evaluation of Iodine-Based Contrast Agents for Ex Vivo Micro-CT Imaging of Porcine Hearts: An Application to Cardiac Conduction System ImagingbioRxiv · 2026
- Role of [123I]I-mIBG cardiac scintigraphy in the differential diagnosis of Parkinson's disease - our experience.Nuclear medicine review. Central & Eastern Europe · 2026
- Variability and bias in the calculation of iodine uptake with an adaptive background subtraction method in contrast-enhanced mammography.Physics in medicine and biology · 2026
- Cardiac sympathetic innervation with [123I]-MIBG SPECT/CT for early prediction of treatment response in wild-type transthyretin amyloidosis: The I-NERVE studyEuropean Heart Journal · 2026
- Radionuclide Dependence of Distance-Dependent Spatial Resolution and Resolution Recovery in SPECT: A SIMIND Monte Carlo Simulation Study.Biomedical physics & engineering express · 2026