Annals of nuclear medicine

Validation of convenient correction methods for collimator-dependent differences in the heart-to-mediastinum ratio from cardiac 123I-MIBG imaging.

Yusuke Inoue, Yutaka Abe, Kohei Mitsui, Hiroki Miyatake, Keita Takahashi, Kei Kikuchi, Eiji Kitamura, Sota Oguro,

Department of Radiological Sciences, Faculty of Health Sciences, Komazawa University, Tokyo, 154- 8525, Japan. inoueys34@gmail.com. · Department of Radiology, Kitasato University Hospital, Sagamihara, Japan.

Key points

From the paper's conclusion
  • Convenient correction methods using 123I point-source imaging were valid for reducing collimator-dependent differences in the H/M ratio in cardiac 123I-MIBG imaging, demonstrating their applicability to a GE system.

Abstract

OBJECTIVE: High-energy photons emitted by 123I may penetrate the septa of a collimator, thereby degrading quantitative accuracy. Collimator choice affects the evaluation of the heart-to-mediastinum (H/M) ratio in cardiac sympathetic imaging with 123I-metaiodobenzylguanidine (123I-MIBG). We investigated the validity of previously reported convenient methods for correcting collimator-dependent differences in the H/M ratio. METHODS: Twenty patients who underwent cardiac 123I-MIBG imaging for evaluation of neurodegenerative diseases on a gamma camera system provided by GE Healthcare were studied. Anterior chest images were acquired successively with three collimators: the low-energy high-resolution and sensitivity (LEHRS), extended low-energy general-purpose (ELEGP), and medium-energy general-purpose (MEGP) collimators. The H/M ratios were estimated using various correction methods: 123I-dual-window (IDW) correction using the high-energy subwindow, empiric correction based on 123I point-source imaging, combined correction using the high-energy subwindow and point-source imaging, and phantom-based method using imaging of a dedicated phantom. Empiric correction and combined correction had been established using patient data mainly obtained with a system provided by Siemens Healthineers. RESULTS: The uncorrected H/M ratios obtained with the LEHRS collimator correlated closely with those obtained with the MEGP collimator but were substantially smaller, indicating severe underestimation with the LEHRS collimator, attributable to septal penetration. The underestimation remained but decreased after any correction. The H/M ratios after correction were largest with combined correction, followed by empiric correction, phantom-based correction, and IDW correction. The absolute errors after correction were smallest for combined correction, followed by empiric correction, phantom-based correction, and IDW correction. The underestimation was substantially smaller with the ELEGP collimator than with the LEHRS collimator, and any type of correction reduced the underestimation. CONCLUSIONS: Convenient correction methods using 123I point-source imaging were valid for reducing collimator-dependent differences in the H/M ratio in cardiac 123I-MIBG imaging, demonstrating their applicability to a GE system.

Read the paper