Real-World safety, patient-reported outcomes, and glycemic metrics with a 7-Day extended infusion set in adults with Type 1 diabetes using automated insulin delivery systems.
María de la O Nieto de la Marca, Pablo Fernández Velasco, Sofía Del Amo Simón, Marta Sánchez Ibáñez, Ana Collantes Matallana, Emilia Gómez Hoyos, Beatriz Torres Torres, Ana Ortolá Buigues,
Centro de Investigación de Endocrinología y Nutrición. Faculty of Medicine, University of Valladolid, Valladolid. Spain, Valladolid, Spain. monieto@uva.es. · Centro de Investigación de Endocrinología y Nutrición. Faculty of Medicine, University of Valladolid, Valladolid. Spain, Valladolid, Spain.
Key points
From the paper's conclusion- In real-world use with an AID system, the E-iS significantly prolonged catheter wear duration and reservoir change intervals while remaining safe and well tolerated. It was associated with high satisfaction and stable glycemic outcomes, supporting E-iS as a practical strategy to reduce the burden of infusion set changes in routine clinical practice.
Abstract
BACKGROUND: Extended-wear infusion sets (E-iS) allowing up to 7 days of catheter use have recently become available. Real-world evidence of contemporary automated insulin delivery (AID) systems using E-iS remains limited. METHOD: A prospective non-randomized single-arm real-world study in adults with type 1 diabetes using the MiniMed-780G AID system who transitioned from conventional 3-day infusion sets to E-iS was conducted. Clinical variables, device settings, adverse events, and patient-reported outcomes were assessed at baseline, 3 and 6 months. RESULTS: 65 participants were included, 58 completed 6 months of follow-up.Mean catheter wear time increased from 3.6 ± 1.1 to 6.1 ± 2.3 days,(p < 0.001), and reservoir change interval increased from 3.5 ± 1.0 to 5.2 ± 2.3 days,(p < 0.001). Glycemic control remained overall stable, although small changes were observed in some CGM metrics (Time in range-TIR: 80.0 ± 7.1% to 78.0 ± 10.0%, p = 0.014) and slight increases in time above range, while hypoglycemia exposure remained unchanged. HbA1c decreased from 6.8 ± 0.5% to 6.4 ± 1.0%, (p = 0.03). Total daily insulin dose did not change, but insulin use related to infusion set management decreased significantly. Pain at cannula insertion improved from 2.1 ± 1.5 to 1.2 ± 1.1,(p < 0.001), and treatment satisfaction increased from 8.2 ± 1.9 to 9.0 ± 1.2,(p = 0.012). No episodes of severe hypoglycemia or diabetic ketoacidosis occurred. Cannula wear duration was strongly correlated with reservoir change interval (r = 0.712, p < 0.001). At 6 months, 92.3% of participants continued using the E-iS, with 48.3% maintaining the full 7-day wear duration. CONCLUSIONS: In real-world use with an AID system, the E-iS significantly prolonged catheter wear duration and reservoir change intervals while remaining safe and well tolerated. It was associated with high satisfaction and stable glycemic outcomes, supporting E-iS as a practical strategy to reduce the burden of infusion set changes in routine clinical practice.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Impact of the use of Extended™ Wear Infusion sets on metabolic, ultrasound, and psychological outcomes in adults with type 1 diabetes using the Medtronic MiniMed 780G: a pilot randomized controlled trial.Endokrynologia Polska · 2026
- A nationwide 12‐month observatory of automated insulin delivery shows improved glucose control, sustained adoption, and reduced acute severe eventsDiabetes Obesity and Metabolism · 2025
- Sustained Improvement in Time in Range After Automated Insulin Delivery Initiation in Adolescents and Adults With Type 1 Diabetes Mellitus: A 12-Month Prospective Observational Study From Saudi Arabia.Journal of diabetes science and technology · 2026
- Extended-Wear Insulin Infusion Sets in People with Type 1 Diabetes: A Scoping ReviewDiabetology · 2026
- Real-world Glycemic Outcomes of a Tubeless Automated Insulin Delivery System (Medtrum A8 TouchCare) in Adults With Type 1 DiabetesCanadian Journal of Diabetes · 2026
- Early Use of Automated Insulin Delivery With Telehealth Support in People Recently Diagnosed With Type 1 Diabetes: One-Year Outcomes From a Prospective Observational Study.Diabetes, obesity & metabolism · 2026