  • We retrospectively reviewed 85 patients (103 knees) from 264 consecutive primary TKAs performed with the imageless hand-held robot (CORI) system at a single center between September 2024 and August 2025.

Abstract

Evidence on hybrid robotic-resection workflows in robotic-assisted TKA is limited. This study quantified agreement between planned, intraoperatively achieved, and early postoperative standing hip-knee-ankle (HKA) alignment using a hybrid resection workflow with the CORI system. We retrospectively reviewed 85 patients (103 knees) from 264 consecutive primary TKAs performed with the imageless hand-held robot (CORI) system at a single center between September 2024 and August 2025. The distal femur was resected and the proximal tibia refined with the burr, and the remaining femoral cuts were made with a conventional 4-in-1 jig. HKA was measured preoperatively, as planned, and as achieved intraoperatively after resections with trials and on full-length weight-bearing radiographs at 3 weeks. Forgotten joint score (FJS 12) was recorded at 6 months. Agreement was assessed using mean signed error, RMSE, and 95% confidence intervals, and measurement reliability using ICC for inter and intra-observer variations. The cohort comprised 85 patients (103 knees; 67 unilateral, 18 bilateral). Mean preoperative HKA was 166.1° ± 6.7° and improved to 177.8° ± 1.9° on early postoperative standing radiographs. Planned HKA was 178.8° ± 0.8° and intraoperatively achieved HKA was 178.0° ± 1.6°, a mean signed error of - 0.8° ± 1.3° (95% CI), indicating slight under-correction relative to plan. RMSE was 1.50° (planned vs achieved), 2.17° (planned vs postoperative), and 1.98° (achieved vs postoperative). Mean FJS 12 at 6 months was 76.0 ± 11.8 (range 54-95) Measurement reliability (inter and intra observer) was good to excellent (ICC 0.89-0.98).In this retrospective series, the hybrid CORI workflow showed close agreement between planned, intraoperatively achieved, and early postoperative standing HKA, with a small mean under-correction of 0.8° relative to plan. These findings are descriptive; without a comparator or long-term clinical outcomes, they cannot establish equivalence to fully robotic or conventional TKA. Prospective comparative studies are needed to evaluate the comparisons.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers