Journal of robotic surgery

Pin-related complications in robotic-assisted total knee arthroplasty: a systematic review and meta-analysis of 15,874 knees.

Michael Ade-Conde, Dheeraj Dhanoa, Jesse I Wolfstadt, Brent A Lanting, Hassaan Abdel Khalik

School of Medicine, Royal College of Surgeons in Ireland, 123 St. Stephen's Green, Dublin, D02 YN77, Ireland. michaeladeconde@gmail.com. · School of Medicine, Royal College of Surgeons in Ireland, 123 St. Stephen's Green, Dublin, D02 YN77, Ireland.

Meta-analysis

Key points

From the paper's conclusion
  • Meta-analysis of four studies (5,635 knees) found no significant difference between intra-incisional pin placement (risk ratio [RR] 0.25, 95% CI 0.03-2.06; P = 0.20; I² = 70%).

Abstract

This review evaluated the incidence, characteristics, and management of pin-related complications following robotic-assisted total knee arthroplasty (RA-TKA). MEDLINE, Embase, and CENTRAL were searched from January 1, 2010, to April 15, 2026 for studies evaluating pin-related complications following primary RA-TKA. Pooled complication rates were estimated using a random-effects generalized linear mixed model, and a random-effects Mantel-Haenszel meta-analysis compared complication rates between intra-incisional and extra-incisional tracker pin placement. Descriptive analyses summarized pin techniques, complications, and management. Thirty-seven studies comprising 15,874 RA-TKAs were included with a mean age of 65.5 years, body mass index of 30.5 kg/m², and follow-up of 10.7 months. Femoral pins were most commonly placed intra-incisionally (54.9%), whereas tibial pins were most commonly placed extra-incisionally (52.5%). The random-effects pooled pin-related complication rate was 0.47% (95% confidence interval [CI] 0.25-0.90%), compared with a crude rate of 0.89% (142/15,874). Wound complications (n = 93, 0.59%) and fractures (n = 24, 0.15%) were most common. Of the 93 wound complications, 69 (74.2%) occurred at extra-incisional tibial sites. Of 40 pin-related complications with reported management, 70% were managed nonoperatively. Meta-analysis of four studies (5,635 knees) found no significant difference between intra-incisional pin placement (risk ratio [RR] 0.25, 95% CI 0.03-2.06; P = 0.20; I² = 70%). Among RA-TKA pin-related complications with reported management, most were managed nonoperatively. Wound complications were most frequently reported at extra-incisional tibial sites, but comparative evidence did not demonstrate a significant difference between placement strategies. Further comparative studies and standardized reporting are needed to optimize pin safety. PROSPERO ID: CRD420261371245. LOE: IV.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers