Robotic-assisted total knee arthroplasty: accuracy, clinical outcomes, complications, learning curve and cost-effectiveness: a critical clinical review.
Jakub Szymczak, Marcin Łapiński, Artur Stolarczyk, Krzysztof Balawender
Department of Orthopaedics and Traumatology, Ostrów Mazowiecka District Hospital, Ostrów Mazowiecka, Poland. jakubszymczak92@gmail.com. · Department of Orthopaedics and Rehabilitation, Medical University of Warsaw, Warsaw, Poland.
Key points
From the paper's conclusion- Its clinical value depends on patient selection, platform, alignment and balancing strategy, surgeon and team experience, volume, and cost structure.
Abstract
Robotic-assisted total knee arthroplasty (RA-TKA) has been adopted to improve surgical planning, bone preparation, component positioning, and limb alignment. This critical clinical review evaluates whether these technical gains improve outcomes that matter to patients and healthcare systems. The PubMed/MEDLINE, Web of Science, and Scopus databases were searched on July 6, 2026, using prespecified RA-TKA search terms and targeted searches for specific endpoints. A supplementary PubMed search on August 10, 2026, focused on soft-tissue balancing, ligament balancing, gap assessment, medial release, and alignment strategy. Evidence was prioritized from systematic reviews, meta-analyses, randomized trials, registry and database studies, health economic evaluations, and learning-curve studies. The most consistent finding is improved radiographic or technical accuracy, including fewer alignment outliers and closer execution of planned implant position. Evidence for patient-reported outcomes, pain, satisfaction, range of motion, and early recovery is more heterogeneous, and reported advantages are often modest or short-term. Safety appears acceptable in trained teams, although pin-related, registration, workflow, and conversion risks require explicit attention. Long-term survivorship superiority and broad cost-effectiveness remain insufficiently proven. RA-TKA improves precision more reliably than it improves durable patient-level outcomes. Its clinical value depends on patient selection, platform, alignment and balancing strategy, surgeon and team experience, volume, and cost structure.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Robotic-assisted total knee arthroplasty: a narrative review of resection accuracy and individualized soft-tissue balance.Journal of robotic surgery · 2026
- Robotic-assisted total knee arthroplasty: radiological precision, clinical outcomes, and benchmarks for emerging systems.Journal of robotic surgery · 2026
- Improved surgical accuracy in total knee arthroplasty using the ROSA® knee system: A randomised‐controlled unblinded trialKnee Surgery Sports Traumatology Arthroscopy · 2025
- Radiographic Accuracy and Consistency of Implant Placement Using the ROSA® Knee System in Robotic-Assisted Total Knee Arthroplasty: A Prospective AnalysisLife · 2026
- Robotic-assisted total knee arthroplasty demonstrates superior early functional outcomes compared with conventional total knee arthroplasty: a quasi-randomised controlled trial.Journal of orthopaedics and traumatology : official journal of the Italian Society of Orthopaedics and Traumatology · 2026
- Robotic arthroplasty and the gap between technical precision and clinical benefit.Journal of robotic surgery · 2026