Determinants of fluoroscopy frequency and effect of standardized workflow optimization in robot-assisted pelvic channel screw placement: a single-center retrospective cohort study.
Jidong Li, Wenhui Zhang, Tingting Gong, Xiaoai Li, Kejun Yin, Xia Zhang, Huiping Tai
Gansu Provincial Hospital, Lanzhou, 730000, China. · Gansu Provincial Cancer Hospital, Gansu Provincial Academic Institute for Medical Research, Lanzhou, 730000, China.
Key points
From the paper's conclusion- We retrospectively enrolled patients with pelvic fractures who underwent robot-assisted percutaneous pelvic channel screw placement at Gansu Provincial Hospital between January 2020 and May 2025.
Abstract
Robot-assisted percutaneous pelvic channel screw placement has been shown to significantly reduce intraoperative radiation exposure. However, even within the robot-assisted framework, fluoroscopy frequency still shows substantial within-group variation. Existing studies lack a systematic analysis of the determinants of fluoroscopy frequency and lack targeted workflow optimization strategies. To identify the independent determinants of fluoroscopy frequency in robot-assisted pelvic channel screw placement and to evaluate whether a standardized optimization protocol based on these factors can further reduce residual fluoroscopy exposure. We retrospectively enrolled patients with pelvic fractures who underwent robot-assisted percutaneous pelvic channel screw placement at Gansu Provincial Hospital between January 2020 and May 2025. Part A: The number of fluoroscopy sessions per case was the dependent variable. Multivariable negative binomial regression was used to analyze potential determinants, including surgeon experience stage, screw type, oblique bone surface at the entry point, BMI, and bone mineral density. Part B: Based on Part A results, we developed and implemented a standardized optimization protocol targeting modifiable factors, with December 2023 as the cutoff. We compared fluoroscopy frequency, fluoroscopy time, and screw accuracy between the pre-optimization group (n = 312) and the post-optimization group (n = 289). To address methodological time-trend confounding, we added interrupted time series analysis, propensity score matching, and a negative control subgroup analysis. This was a non-randomized, single-center, before-after design, and residual confounding cannot be completely excluded. Part A (n = 601): Multivariable negative binomial regression showed that oblique bone surface at the entry point (IRR = 1.52, 95% CI: 1.31-1.77, P < 0.001), early surgeon experience stage (first 10 cases vs. > 30 cases, IRR = 1.48, 95% CI: 1.25-1.75, P < 0.001), S2 screws (IRR = 1.26, 95% CI: 1.10-1.44, P = 0.001), and BMI ≥ 30 kg/m² (IRR = 1.19, 95% CI: 1.02-1.39, P = 0.026) were independent determinants of fluoroscopy frequency. Part B: The post-optimization group showed a significant reduction in fluoroscopy frequency compared with the pre-optimization group (9.8 ± 2.9 vs. 13.5 ± 3.7, P < 0.001). Fluoroscopy time also decreased (16.2 ± 5.4 s vs. 21.8 ± 6.9 s, P < 0.001). Screw accuracy showed no significant difference (98.7% vs. 98.1%, P = 0.52). The optimization effect was most pronounced in the oblique bone surface subgroup (31.5% reduction) and was 14.2% in the non-oblique subgroup. In the negative control subgroup (experienced surgeons, non-oblique, non-obese, S1 screws), fluoroscopy frequency did not change significantly before and after optimization (10.2 ± 2.6 vs. 10.5 ± 2.8, P = 0.41); however, post-hoc power for this subgroup was extremely low (approximately 0.10), so this null result should be regarded as inconclusive rather than as evidence of no effect. These findings suggest a possible risk-stratified specific effect of the optimization protocol, but time trends, the Hawthorne effect, and the unseparated multicomponent bundle cannot be excluded. In robot-assisted pelvic channel screw placement, oblique bone surface at the entry point, early surgeon experience stage, S2 screws, and obesity are independent determinants of fluoroscopy frequency. A standardized multicomponent optimization protocol targeting modifiable factors can safely reduce fluoroscopy frequency by approximately 27% without compromising screw accuracy. Optimization of fluoroscopy frequency should become a core indicator in robot-assisted surgical training and quality control. However, this was a non-randomized, single-center, before-after study; the optimization protocol was a multicomponent bundle whose individual contributions cannot be separated; and the oblique bone surface and negative control subgroups were underpowered. These findings require prospective validation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Axial Convergence in Percutaneous Robotic Pedicle Screw Placement: Mixed-Effects Analysis of 1006 ScrewsSpine Open · 2026
- Robot-assisted versus freehand cannulated-screw fixation for femoral neck fractures: a systematic review of technical, clinical and adoption outcomes.Journal of robotic surgery · 2026
- Robot-assisted lumbar facet joint infiltration improves accuracy and reduces radiation exposure compared to the manual technique in a comparative phantom studyScientific Reports · 2026
- Robot-assisted versus pedicle projection localization in unilateral biportal endoscopic spine surgery: a trade-off analysis of fluoroscopy use, procedural time, and learning curve.Journal of robotic surgery · 2026
- Clinical outcomes and learning curve of robot-assisted percutaneous cannulated screw fixation for femoral neck fractures.Frontiers in surgery · 2026
- Comparative evaluation of CT-to-Fluoroscopy and Scan-and-Plan registration techniques in robot-assisted pedicle screw placement: a preliminary cadaveric faesibility study.Journal of robotic surgery · 2026