Intraoperative Impedance Planimetry Reveals Functional Sleeve Geometry Beyond Fixed Bougie Calibration.

Michael de Cillia, Jonas Bolm, Martin Grünbart, Michael Weitzendorfer, Hannes Hoi, Alexandra Majerus, Robert Uzel, Pia Tscholl,

Department of Surgery, Hospital of the Brothers of St. John of God, Salzburg, Austria. michael.decillia@bbsalz.at. · Department of Surgery, Hospital of the Brothers of St. John of God, Salzburg, Austria.

Observational

Key points

From the paper's conclusion
  • Despite standardized SG with a fixed bougie size, substantial variability persisted in luminal EndoFLIP™ parameters, indicating that bougie diameter alone does not fully determine the measured geometry of the newly constructed sleeve. Intraoperative impedance planimetry identified Dmin and CSA as exploratory candidate markers associated with postoperative nausea during index hospital stay. Further prospective studies are required to assess its value as a step towards physiology-guided SG.

Abstract

BACKGROUND: Rigid bougies are commonly used for intraoperative sleeve calibration in standardized sleeve gastrectomy to ensure a reproducible sleeve geometry. Current evidence on bougie size and its impact on postoperative outcome and sleeve shape remains inconclusive and inconsistent. This pilot study was conducted to assess whether a fixed bougie size leads to reproducible sleeve configuration and to evaluate the relationship between early postoperative outcome and objective intraoperative luminal sleeve geometry measurements. METHODS: A prospective analysis was performed in patients providing informed consent for sleeve gastrectomy (SG). SG using a 32Fr bougie and the intraoperative measurements followed a meticulous standardized protocol. Distensibility index (DI), cross-sectional area (CSA), minimal diameter (Dmin), high-pressure zone (HPZ) and intra-balloon pressure (IBP) served as objective measurement parameters of functional sleeve geometry, and correlation analyses with postoperative outcomes were performed. RESULTS: SG with objective intraoperative sleeve evaluation was technically feasible in all patients and no device-related adverse events were observed in this pilot cohort. Luminal measurements showed exploratory associations with early clinical outcome, with lower Dmin and CSA associated with a higher occurrence of postoperative nausea (Dmin: 95% CI -1.00 to -0.357; CSA: 95% CI -0.947 to -0.188], respectively). Exploratory ROC analysis identified cohort-derived candidate cut-off values of 18.15 mm for Dmin and 259.5 mm² for CSA. CONCLUSION: Despite standardized SG with a fixed bougie size, substantial variability persisted in luminal EndoFLIP™ parameters, indicating that bougie diameter alone does not fully determine the measured geometry of the newly constructed sleeve. Intraoperative impedance planimetry identified Dmin and CSA as exploratory candidate markers associated with postoperative nausea during index hospital stay. Further prospective studies are required to assess its value as a step towards physiology-guided SG.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Intraoperative Impedance Planimetry Reveals Functional Sleeve Geometry Beyond Fixed Bougie Calibration — Science Skim