Early-phase single-time-point kidney dosimetry in 177Lu-DOTATATE therapy using a half-life fraction-based correction: a retrospective single-center evaluation.

Anji Yokouchi, Takayuki Shibutani, Takahiro Konishi, Hiroto Yoneyama, Kosuke Matsubara, Hiroshi Wakabayashi

Department of Radiological Technology, Kanazawa University Hospital, Kanazawa, Japan. · Division of Health Sciences, Department of Quantum Medical Technology, Graduate School of Medical Sciences, Kanazawa University, Kanazawa, Japan. iwsb03100621@staff.kanazawa-u.ac.jp.

Observational · n=17

Key points

From the paper's conclusion
  • HLF-corrected STP dosimetry is feasible during short-stay PRRT for toxicity-oriented renal dose estimation. STP (4 h) showed no systematic bias against the reference. STP (24 h) systematically overestimated it but fits routine radiation-safety and scheduling constraints. Both findings require confirmation in independent cohorts. Neither approach replaces multi-time-point dosimetry.

Abstract

PURPOSE: Multi-time-point kidney dosimetry for 177Lu-DOTATATE peptide receptor radionuclide therapy (PRRT) is resource-intensive, and delayed-phase imaging is infeasible where PRRT is delivered during a short hospitalization. We evaluated a single-time-point (STP) method using a cohort-specific half-life fraction (HLF) correction at 4-120 h post-injection. MATERIALS AND METHODS: We retrospectively analyzed 17 patients (34 kidneys) after a first 177Lu-DOTATATE cycle (October 2021-January 2024). Reference doses came from a 4-time-point (4TP) SPECT/CT protocol at 4, 24, 72 and 120 h. STP doses were compared with the reference by mean percent error (MPE), root mean square error (RMSE), Bland-Altman analysis and the Friedman test with Holm post-hoc correction. Four-fold cross-validation, with the HLF re-derived from training folds only, was applied at 4 and 24 h. RESULTS: The mean 4TP dose was 3120.5 ± 1015.7 mGy. STP (4 h) gave 3307.6 ± 1131.7 mGy (not significantly different), whereas 24, 72 and 120 h significantly overestimated it. The whole-cohort MPE was 8.3 ± 31.1% for STP (4 h) and 27.9 ± 26.0% for STP (24 h). The Bland-Altman bias of STP (4 h) (4TP - STP, -187 mGy) did not differ from zero, whereas that of STP (24 h) (-​​818 mGy) did. Cross-validated errors (MPE 17.1 ± 14.6% at 4 h, 9.3 ± 8.8% at 24 h) may have been affec​ted by differences involumes of interest (VOIs) ​between analyses and did not establish the superiority of either time point. CONCLUSION: HLF-corrected STP dosimetry is feasible during short-stay PRRT for toxicity-oriented renal dose estimation. STP (4 h) showed no systematic bias against the reference. STP (24 h) systematically overestimated it but fits routine radiation-safety and scheduling constraints. Both findings require confirmation in independent cohorts. Neither approach replaces multi-time-point dosimetry.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers