Development and implementation of an automated micrometric bone distraction system based in a fuzzy logic controller for an Aybar-type external fixator.
Joao Marcelo Valdivia De La Gala, Andrea Sophia Valladares Chavez, Carlos Hernan Inga Espinoza
School of Mechatronics Engineering, Universidad Peruana de Ciencias Aplicadas, Lima, 15088, Lima, Perú. u202018990@upc.edu.pe. · School of Mechatronics Engineering, Universidad Peruana de Ciencias Aplicadas, Lima, 15088, Lima, Perú.
Key points
From the paper's conclusion- These findings demonstrate that it is possible to enhance the precision and reproducibility of the Aybar external fixator through low cost automation.
Abstract
Bone distraction is a widely applied orthopedic procedure based on Ilizarov's tension-stress principle, which requires high precision and accuracy to guarantee successful bone regeneration. In many low resource locations, the Aybar-type external fixator is the only viable treatment option due to its low cost and accessibility. However, its manual procedure introduces human error, reduces precision, and compromises treatment outcomes. To address this problem, this study develops and implements an automated micrometric distraction system for an Aybar-type fixator, aiming to provide a low cost yet highly accurate alternative, particularly focused on pediatric patients, for whom precision is critical to ensure proper callus formation and bone growth. Four control strategies were designed and evaluated: an ON-OFF controller, a Proportional-Integral (PI) controller, a Linear Quadratic Integrator (LQI), and a Fuzzy Logic Controller (FLC). The mechanical platform integrates a Direct Current (DC) motor, encoder, and worm gear with 1.25 mm pitch, while simulations and experimental tests were performed under two distraction regimens: 1 mm/day in four increments and 1 mm/day in sixty increments. Results indicate that the FLC achieves the best performance in coarse step elongation, while the LQI provides superior results in high-resolution protocols. These findings demonstrate that it is possible to enhance the precision and reproducibility of the Aybar external fixator through low cost automation. The proposed solution contributes to improving treatment quality for pediatric patients in contexts where advanced imported systems are unavailable, filling a critical gap between affordability and clinical efficacy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A model predictive displacement controller with an adaptive ESO and fuzzy dead-zone compensator of drill-arm for a hydraulic roofbolter.ISA transactions · 2026
- Temperature Increase Limitation During Automatic Bone Drilling in Orthopedic Surgery.International journal of medical sciences · 2026
- Γ-Cantilever Mechanism-Informed BP Neural Network Compensation for Gap Measurement of an Integrated Knee Distractor in Total Knee Arthroplasty.Sensors (Basel, Switzerland) · 2026
- Evaluation of a Fuzzy-Supervised PID Controller for a Parallel Rehabilitation Mechanism.Bioengineering (Basel, Switzerland) · 2026
- Enhanced Bone Healing in Tibial Osteogenesis Using the Distraction Holiday Protocol in Ilizarov Fixation.Indian journal of orthopaedics · 2026
- Finite element evaluation of a novel adjustable-length partial ossicular replacement prosthesis.Frontiers in bioengineering and biotechnology · 2026