Veterinary research communications

Comparison of 30° and 70° arthroscopes in canine hip arthroscopy: a 3D-printed model study.

Piotr Trębacz, Mateusz Pawlik, Michał Czopowicz, Anna Barteczko, Aleksandra Kurkowska, Agata Piątek

Department of Small Animal Surgery and Anaesthesiology, Institute of Veterinary Medicine, Warsaw University of Life Sciences-SGGW, Nowoursynowska 159 C, Warsaw, 02-776, Poland. piotr_trebacz@sggw.edu.pl. · Department of Biomaterials and Medical Devices Engineering, Faculty of Biomedical Engineering, Silesian University of Technology, Zabrze, Poland.

Key points

From the paper's conclusion
  • The results suggest that, in this experimental model, the 70° arthroscope provided a comparable overall visible area of the artificial hip joint surface to the 30° arthroscope during dorso-lateral canine hip arthroscopy.

Abstract

Hip arthroscopy is a valuable diagnostic and surgical tool in veterinary medicine; however, the optimal arthroscope viewing angle for the canine coxofemoral joint has not been established. We hypothesised that a 70° arthroscope would provide superior visualisation of the hip joint cavity compared to the standard 30° arthroscope when inserted via a dorso-lateral approach. To test this hypothesis, five 3D-printed canine hip joint models were developed using computed tomography (CT) scans of large-breed dogs. The articular surfaces were embossed with a 1 × 1 mm checkerboard pattern to enable objective quantification of the visible area. Arthroscopy was performed with 2.7-mm-diameter 30° and 70° arthroscopes at four positions (3, 6, 9, and 12 o'clock), and the visible area was calculated by counting visible tiles. Statistical analysis was performed using a mixed-effect general linear model. There was no significant overall difference in the visible area between the two arthroscope viewing angles (P = 0.977). However, arthroscope position significantly affected the visible area (P < 0.001), and a significant interaction between viewing angle and position was identified (P = 0.001). Notably, the visible area produced by the 70° arthroscope was significantly smaller than that produced by the 30° arthroscope at the 12 o'clock position (P = 0.026). However, at the 3 o'clock position, the visible area produced by the 70° arthroscope appeared to be larger than that produced by the 30° arthroscope. Nevertheless, this difference was not significant (P = 0.085). The results suggest that, in this experimental model, the 70° arthroscope provided a comparable overall visible area of the artificial hip joint surface to the 30° arthroscope during dorso-lateral canine hip arthroscopy. Further comparative studies, particularly on calibrated cadaveric hip joints, are required.

Read the paper