科研速览 · Science Skim继续刷下去 · Keep skimming →
◆ Frontiers in public health2026-01-01

Superior-head automated external defibrillator placement during single-rescuer cardiopulmonary resuscitation: a randomized crossover simulation study and retrospective multicenter cohort.

Daojian Xu, Jingchen Zhang, Honghang Lin, Qiqi Cai, Sheng Zhang, Jueyue Yan

一句话结论 · In one sentence

Superior-head AED placement improved the efficiency, compression continuity, and workload of simulated single-rescuer CPR. The retrospective clinical findings were directionally consistent but statistically inconclusive and should be regarded as hypothesis-generating rather than confirmatory. Prospective validation is required before clinical implementation.

原始摘要(英文原文)· Original abstract
BACKGROUND: During single-rescuer cardiopulmonary resuscitation (CPR), one rescuer must sustain chest compressions while operating an automated external defibrillator (AED), yet the optimal physical position of the AED within the rescuer's working area is uncertain. The study evaluated whether superior-head AED placement improves the single-rescuer defibrillation workflow. METHODS: The study combined a randomized four-sequence crossover manikin simulation (32 healthcare providers; 128 evaluable trials) comparing four prespecified AED positions-superior-head placement (position D) versus alternative placements (positions A-C)-with a retrospective cohort of 213 AED-treated adult cardiac-arrest cases from six centers (57 with position D, 156 with other positions). The primary simulation outcome was compression-to-first-shock time; the primary clinical outcome was the collapse-to-first-shock interval. Secondary outcomes included AED-related hands-off time, chest compression fraction, NASA Task Load Index workload, and procedural safety. RESULTS: In the simulation, position D shortened compression-to-first-shock time by 14.8-26.3 s relative to the other positions (all Holm-adjusted p < 0.001), reduced AED-related hands-off time, increased chest compression fraction (83% vs. 78-80%), and lowered perceived workload (all p < 0.001), without increasing critical errors; over-body actions were least frequent with position D. In the retrospective cohort, the adjusted median difference in the collapse-to-first-shock interval was -0.42 min (approximately 25 s; 95% CI, -0.91 to 0.07; p = 0.092); estimates consistently favored position D across sensitivity, restricted-cohort, weighted, and leave-one-center-out analyses, although every confidence interval crossed the null. Patient outcomes did not differ by AED position. CONCLUSION: Superior-head AED placement improved the efficiency, compression continuity, and workload of simulated single-rescuer CPR. The retrospective clinical findings were directionally consistent but statistically inconclusive and should be regarded as hypothesis-generating rather than confirmatory. Prospective validation is required before clinical implementation.
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

Superior-head automated external defibrillator placement during single-rescuer cardiopulmonary resuscitation: a randomized crossover simulation study and retrospective multicenter cohort. — 科研速览 Science Skim