An Optomechanical Analysis Revealing the Importance of Haptic Design and Material Selection in Shaping the Performance of Dual-Optic Accommodative IOLs.
Yunxin Xia, Jiangzhen Guo, Chunjing Tao, Barbara K Pierscionek, Kehao Wang
School of Engineering Medicine, Beijing Advanced Innovation Center for Biomedical Engineering, Beihang University, No.37 Xueyuan Road, Haidian District, Beijing, 100191, China. · School of Engineering Medicine, Beijing Advanced Innovation Center for Biomedical Engineering, Beihang University, No.37 Xueyuan Road, Haidian District, Beijing, 100191, China. jzguo@buaa.edu.cn.
Key points
From the paper's conclusion- Under the same applied radial displacement, the multi-folded haptic design provides greater power change per unit force than the mono-folded and non-folded designs. The simulated paraxial power response and biomechanical behavior of the dual-optic AIOL are jointly determined by haptic design and material selection, which should be considered in future AIOL development.
Abstract
PURPOSE: Accommodative Intraocular Lenses (AIOLs) have long been thought to be a promising solution for restoring visual accommodation after cataract surgery, but their performance critically depends on both the geometric design and the selected materials. The present study aims to compare how haptic design and material combination affect the simulated paraxial power response and mechanical behavior of selected dual-optic AIOL configurations under prescribed radial compression. METHODS: Twelve dual-optic AIOL finite element models were constructed by combining three different haptic designs (non-folded, mono-folded, and multi-folded) with four different combinations of materials. A prescribed radial inward displacement was applied to each model to analyze the deformation, reaction force, and stress distributions. The resulting deformations were then used for modeling optical ray tracing to evaluate optical power changes. Comparative analyses were conducted to assess the influence of haptic design and material parameters on the paraxial power response and mechanical performance of the AIOLs. RESULTS: Both the haptic design and optic materials affect the optomechanical behavior of the AIOL. At a common radial displacement, multi-folded and mono-folded haptics yield greater power change than non-folded haptics. The multi-folded design yields greater power change per unit force than the mono-folded design. Material combinations also influence refractive power changes. High stresses were concentrated at the peripheral region of the anterior optic for mono- and multi-folded designs and at the junction between posterior optic and haptic for the non-folded design. CONCLUSION: Under the same applied radial displacement, the multi-folded haptic design provides greater power change per unit force than the mono-folded and non-folded designs. The simulated paraxial power response and biomechanical behavior of the dual-optic AIOL are jointly determined by haptic design and material selection, which should be considered in future AIOL development.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative Two-Year Outcomes of Three Different Three-Piece Intraocular Lenses in Flanged Intrascleral Fixation.Medicina (Kaunas, Lithuania) · 2026
- Long-term visual outcomes and rotational stability of two diffractive toric intraocular lenses with different design combinations in cataract patients with long axial length.Frontiers in medicine · 2026
- Multifunctional Injectable Hydrogel as a Biomimic Lens: Optical–Mechanical Restoration and Dynamic Postoperative Microenvironment RegulationAdvanced Materials · 2026
- Comparative safety and complication profiles of two three-piece intraocular lenses in Yamane technique.International ophthalmology · 2026
- Haptic-Extended Flange Intrascleral Fixation for Single-Piece Acrylic Intraocular Lenses.Journal of clinical medicine · 2026
- Comparative Optical Performance of Two Enhanced Monofocal IOLs Across Pupil Sizes and Lens Powers.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026