Smartphone-Based Cuffless Blood Pressure Estimation Using a Multiple Regression Model: Revalidation and On-Demand Performance Under Real-World Conditions.
Takehiro Yamakoshi, Peter Rolfe, Ken-Ichi Yamakoshi
Institute of Liberal Arts and Science, Kanazawa University, Kakuma-machi, Kanazawa, Ishikawa, 920-1192, Japan. takehiroy@me.com. · Department of Automatic Measurement and Control, Harbin Institute of Technology, No.92 West Dazhi Street, Nan Gang District, Harbin, 150001, Heilongjiang Province, China.
Key points
From the paper's conclusion- The proposed MRM-based method demonstrates practical BP estimation accuracy in daily life. Although further calibration studies are warranted, its long-term stability and smartphone compatibility offer notable advantages for routine BP monitoring. This approach represents a promising, convenient cuffless alternative for integration into mobile healthcare platforms and daily BP management.
Abstract
PURPOSE: Consistent blood pressure (BP) management is crucial for cardiovascular health; yet conventional cuff-based methods remain cumbersome for frequent use. Although cuffless BP technologies are advancing, further validation in daily life is still required. Building on our preliminary work, this study aims to revalidate and further develop a practical, sustainable smartphone-based BP estimation method using multiple regression model (MRM) and assess on-demand performance under real-world conditions. METHODS: Calibration models were constructed using 101 datasets from 13 healthy participants. Target variables were brachial cuff-based systolic and diastolic BP (SBPc/DBPc). Predictor variables reflecting BP regulation were pulse rate index (PRI: ln(PR)) and peripheral vascular resistance index (VRI: |ln(mNPV)|; mNPV: modified normalized pulse volume), derived from smartphone-based fingertip green-light photoplethysmography. Model performance was evaluated by independent validation in additional participants (n = 11) under normal living conditions by comparing estimated BP (SBPe/DBPe) with concurrent cuff-based measurements. RESULTS: Over 1-2 weeks, 168 datasets (13-21 per participant) were collected. Mean absolute errors (MAEs) were 3.2 ± 2.59 mmHg for SBP and 4.7 ± 2.78 mmHg for DBP. A longitudinal case study over 20 months in one participant (752 datasets) demonstrated lower MAEs: 1.6 ± 1.16 mmHg for SBP and 2.4 ± 1.71 mmHg for DBP. CONCLUSION: The proposed MRM-based method demonstrates practical BP estimation accuracy in daily life. Although further calibration studies are warranted, its long-term stability and smartphone compatibility offer notable advantages for routine BP monitoring. This approach represents a promising, convenient cuffless alternative for integration into mobile healthcare platforms and daily BP management.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Lightweight Calibration-Based Blood Pressure Tracking from Photoplethysmography Features for Smart Devices: Accuracy, Model Size and Recalibration Interval.Diagnostics (Basel, Switzerland) · 2026
- Context-Augmented, Session-Calibrated Blood Pressure Tracking from PhotoplethysmographyarXiv · 2026
- Multi-Wavelength PPG Amplitude Ratio Features for Cuffless Blood Pressure Estimation: An Exploratory Study.Sensors (Basel, Switzerland) · 2026
- Validation of a wrist-worn dual-sensor applanation tonometry device for blood pressure measurement under rest and stress conditionsEuropean Heart Journal · 2026
- Continuous Cuffless Blood Pressure Estimation Using PPG From Methods to Clinical TranslationarXiv · 2026
- Wearable, cuffless, and portable devices for blood pressure monitoring (2015-2025): a scoping review of technologies, calibration, clinical validation, and regulatory frameworks.Frontiers in digital health · 2026