Scoliosis Induction in the Porcine Spine Using a Growth-Modulating Posterior Tether: Finite Element Model Development and Calibration.
Christian R D'Andrea, Mattan R Orbach, Alessandra Fusco, Edward J Vresilovic, Brian D Snyder, Thomas P Schaer, Patrick J Cahill, Sriram Balasubramanian
School of Biomedical Engineering, Science, and Health Systems, Drexel University, Philadelphia, PA, 19104, USA. · Sidney Kimmel Medical College, Thomas Jefferson University, Philadelphia, PA, 19107, USA.
Key points
From the paper's conclusion- Image-based FEMs estimated vertebral stress-growth relationships and reproduced deformity with comparable Cobb angle error to an existing parametric FE model (current 2.5 ± 1.6° vs. reported ~ 2.0 ± 2.4° in a single-pig matched comparison) with realistic baseline IVD stresses (current ~ 0.45 MPa vs. reported 0.48 MPa in the growth plate). With expanded validation, such models may support preclinical evaluation of implant designs and surgical configurations prior to in vivo testing.
Abstract
PURPOSE: Predictable correction of early onset scoliosis through mechanical modulation of endochondral growth requires understanding vertebral stress-growth relationships. Here, we develop image-based finite element models (FEMs) to estimate vertebral stress-growth relationships under prescribed loading conditions using a porcine model of spine deformity induced mechanically via a posterolateral tether, assuming that, after initial tether-induced elastic deformation, subsequent deformity develops solely from vertebral endochondral growth against the tether and elastic anatomical structures. METHODS: Image-based FEMs were generated to simulate juvenile Yorkshire pigs instrumented with a posterolateral tether implant which applied asymmetric force across the spine. FEM growth parameters were calibrated to best match in vivo data using landmark points obtained from 3D imaging for each pig. RESULTS: Asymmetric stresses generated by loading of the posterolateral offset cable produced multi-planar spine deformity. Calibrated FEMs achieved moderate agreement with frontal plane projections of the deformity: Cobb angle (MAE = 4.6-8.5° across all pigs and timepoints) and anatomic landmark point correspondence (MAE = 2.1-2.9 mm), but sagittal plane agreement was more limited, partly attributable to variability in positioning during imaging. FEM-derived vertebral body growth rates were reduced for vertebrae along the concavity of the deformity. CONCLUSION: Image-based FEMs estimated vertebral stress-growth relationships and reproduced deformity with comparable Cobb angle error to an existing parametric FE model (current 2.5 ± 1.6° vs. reported ~ 2.0 ± 2.4° in a single-pig matched comparison) with realistic baseline IVD stresses (current ~ 0.45 MPa vs. reported 0.48 MPa in the growth plate). With expanded validation, such models may support preclinical evaluation of implant designs and surgical configurations prior to in vivo testing.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Influence of scoliosis severity on thoracolumbar biomechanical properties: a finite element analysis-based study.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2026
- Biomechanical analysis of T10-T11 vertebrae under the combined effects of osteoporosis and L4 spondylolisthesis: a finite element analysis based on computed tomography.Frontiers in bioengineering and biotechnology · 2026
- Biomechanical effects of traditional Chinese exercises on scoliosis: a combined finite element and surface electromyography analysis.Frontiers in bioengineering and biotechnology · 2026
- Biomechanical comparative analysis of minimally invasive decompression and minimally invasive fusion with internal fixation in the treatment of lumbar spinal stenosis with degenerative spondylolisthesis: a finite element study.BMC musculoskeletal disorders · 2026
- Modeling rod elasto-plasticity improves biomechanical simulation of adolescent idiopathic scoliosis instrumentation.Computer methods in biomechanics and biomedical engineering · 2026
- Direction-dependent biomechanical effects of transverse process hooks versus pedicle screws at the uppermost instrumented vertebra in adult spinal deformity: a finite element analysis.Frontiers in bioengineering and biotechnology · 2026