A Preclinical Approach to 3D-Printed Osseointegration Prosthesis: Improving Functional Outcomes and Reducing Complications in Residual Limb Reconstruction.
Shuaihe Gong, Yifan Wang, Chang Zhou, Jiale Li, Kexin Xie, Wei Guo, Yi Yang, Jun Wang
Department of Musculoskeletal Tumor, Peking University People's Hospital, No. 11 Xizhimen South Street, Xicheng District, Beijing, 100044, China. · Department of Musculoskeletal Tumor, Peking University People's Hospital, No. 11 Xizhimen South Street, Xicheng District, Beijing, 100044, China. pkuwangjun@bjmu.edu.cn.
Key points
From the paper's conclusion- A mid-femoral amputation model was constructed in Small-tailed Han sheep (n = 3), with Type IIa OIP press fitted after intramedullary cavity reaming and cortical decortication, evaluating functional reconstruction, biocompatibility, and osseointegration.
Abstract
This study presented a novel additively manufactured osseointegration prosthesis (OIP) for patients with amputation residual limb functional reconstruction, elaborating its design concept, conducting finite element mechanical analysis, and evaluating efficacy via animal models to provide a clinical theoretical basis. Our designed OIP system were categorized into four types: Type I (solely intramedullary press-fit fixation) and Type II (combined intramedullary press-fit-extramedullary sleeve fixation); Type a (no auxiliary anti-rotation screws) and Type b (with such screws). The cloverleaf sleeve's inner wall and intramedullary stem surface featured 3D-printed metal porous structures. Finite element models were established using adult femoral CT data to simulate stress/displacement under 15,000/20,000 N*mm torque and 600/900 N vertical stress. A mid-femoral amputation model was constructed in Small-tailed Han sheep (n = 3), with Type IIa OIP press fitted after intramedullary cavity reaming and cortical decortication, evaluating functional reconstruction, biocompatibility, and osseointegration. Finite element analysis yielded stress peaks, maximum displacement, and distribution contour plots. Results showed cloverleaf sleeve-equipped prostheses had lower stress peaks under high vertical stress, more dispersed stress distribution, and superior displacement resistance. Animal model data demonstrated stable fixation, no obvious infection, and favorable prosthesis-femur osseointegration. This newly designed OIP possesses good biocompatibility and osseointegration capacity and the extramedullary cloverleaf sleeve enhances mechanical properties, laying a theoretical foundation for future clinical application.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Influence of the pelvic bone and gluteal zone over the stress state at the interaction between the prosthetic socket and residual limb for people with transfemoral amputation.Frontiers in bioengineering and biotechnology · 2026
- Mechanical evaluation of 3D-printed above-knee prosthetic sockets using a biofidelic hyperelastic silicone limb surrogate.Medical engineering & physics · 2026
- A novel strategy for reconstruction of AORI type III proximal tibial defects : surface-type knee reconstruction following defect reconstruction with 3D-printed extension-stem augment.Bone & joint research · 2026
- Computational analysis of Lattice-Integrated transfemoral prosthetic socketsMaterials & Design · 2026
- Modular 3D-printed axis prosthesis for large C2 bone defect reconstruction based on Chinese population anatomical data: finite element analysis and preliminary clinical outcomes.BMC musculoskeletal disorders · 2026
- Modular Cage-Type Prosthetic Construction for Lumbar Intervertebral Reconstruction: A Computer-Aided Design-Based Geometric Design and Finite Element Analysis Study.Cureus · 2026