Ionic characteristics and osmolarity of cooling fluid in hepatic laser ablation: a 7T MRI-based comparison in ex vivo bovine liver.
Fiona Mankertz, Josephine Berger, Nour Maalouf, Benedikt Stenzl, Antonia Ashkar, Ole Gemeinhardt, Johannes Uhlig, Saif Afat,
Department of Diagnostic and Interventional Radiology, Universitätsklinikum Tübingen, Tübingen, Germany. fiona.mankertz@med.uni-tuebingen.de. · Department of Diagnostic and Interventional Radiology, Universitätsklinikum Tübingen, Tübingen, Germany.
Key points
From the paper's conclusion- In ex vivo bovine liver, 10% mannitol was associated with larger MRI-defined laser ablation volumes than 0.9% sodium chloride or 5% dextrose.
Abstract
To compare MRI-defined ablation volume among three cooling fluids with different ionic characteristics and osmolarity in ex vivo bovine liver tissue. In this ex vivo pilot study, 36 laser ablations were performed in five bovine livers using internal applicator cooling with 0.9% sodium chloride, 5% dextrose in water, or 10% mannitol at 60 mL/h. Post-ablation MRI was performed at 7 T. Primary endpoint was manual MRI volumetry; secondary endpoint was maximum axial diameter. Semi-automatic volumetry and intra-rater repeatability were assessed as supplementary analyses. Linear mixed-effects models with source liver as a random intercept were used for group comparisons. Overall, 33/36 ablations were evaluable. Mean ablation volume was 21.4 ± 1.5 cm³ with sodium chloride, 21.7 ± 1.4 cm³ with dextrose, and 23.6 ± 1.3 cm³ with mannitol. Cooling fluid group was associated with ablation volume (p = 0.002). Ablation volume was larger with mannitol than with sodium chloride (mean difference, 2.2 cm³; adjusted p < 0.001) or 5% dextrose (mean difference, 1.9 cm³; adjusted p = 0.006). Maximum axial diameter showed the same group pattern. Agreement between manual and semi-automatic volumetry was high (ICC, 0.96); intra-rater repeatability was excellent for volume (ICC, 0.98) and diameter (ICC, 0.97). In ex vivo bovine liver, 10% mannitol was associated with larger MRI-defined laser ablation volumes than 0.9% sodium chloride or 5% dextrose.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- New analytical and hybrid heat transfer models for thermal ablation procedures validated by MRI thermometryInternational Journal of Hyperthermia · 2025
- Gelified ethanol as a low-cost non-thermal ablative agent for percutaneous liver ablation: a first-phase preclinical feasibility studyCVIR Oncology · 2026
- A viscosity-associated fluorescence turn-on probe for rapid local and topical tumor visualization.Chemical communications (Cambridge, England) · 2026
- Assessment of temperature dynamics in pulsed field ablation with a variable-loop circular catheter: a comparative analysis of waveform configurations and irrigation rates in specimens of bovine ventricular myocardiumEP Europace · 2025
- Imaging Informatics for Thermal Ablation: CT Thermometry-Based Tissue Injury Mapping.IEEE journal of biomedical and health informatics · 2026
- Temperature-dependent nonlinear response in transient radio frequency heating of saline waterAIP Advances · 2026