Static and fatigue behavior of one- and two-piece zirconia implant abutment systems: an in vitro study.
Pedro Diaz, Monica Lopez, Luis J Gil Villagra, Juan Carlos Del Real, Seyed Ali Mosaddad, Maria J Suarez
Department of Conservative Dentistry and Prosthodontics, Faculty of Dentistry, Complutense University of Madrid (UCM), Plaza Ramón y Cajal s/n, Madrid, 28040, Spain. · Department of Clinic Grade Dentistry, Faculty of Odontology, European University of Madrid, Madrid, Spain.
Key points
From the paper's conclusion- Static fracture-load superiority was not demonstrated. Fatigue run-out varied with relative loading; modeled stress differences do not establish intrinsic material-strength differences.
Abstract
OBJECTIVES: To compare static fracture behavior, fatigue survival, and fracture patterns of three one- and two-piece zirconia implant abutment systems. MATERIALS AND METHODS: Three groups comprised 48 assemblies each: (1) Astra Tech ZirDesign, one-piece zirconia; (2) Biomet 3i CAP454, two-piece zirconia with a titanium base; and (3) Straumann RC Anatomic, one-piece zirconia. Within each group, 12 assemblies underwent initial static testing and 12 each underwent fatigue at 25%, 50%, or 75% of the system-specific reference load (10 Hz, approximately 30°, five million cycles). Static outcomes underwent one-way and Welch analyses of variance; pooled run-out proportions underwent exact tests (α = 0.05). RESULTS: Mean static fracture loads were 1058 ± 225 N (Astra Tech), 866 ± 189 N (Biomet 3i), and 873 ± 402 N (Straumann), without a significant difference (p = .194). Calculated tensile-side stress differed (p = .016); compressive-side findings were method-dependent. Run-out was 88.9% at 25% (32/36; 95% confidence interval, 73.9%-96.9%) and 0% at higher levels (p < .001). Residual fracture outcomes described survivors only. Fractography indicated brittle cervical failure. CONCLUSIONS: Static fracture-load superiority was not demonstrated. Fatigue run-out varied with relative loading; modeled stress differences do not establish intrinsic material-strength differences. CLINICAL RELEVANCE: These system-specific laboratory results cannot isolate an independent titanium-base effect or establish clinical service life.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- 3D-printing for zirconia abutment manufacturing and screw access hole effect: fatigue behavior of milled zirconia simplified restorations.Odontology · 2026
- Influence of Abutment Configuration on the Fatigue Survival of Titanium Implant-Abutment Assemblies.Materials (Basel, Switzerland) · 2026
- Clinical Performance and Failure Mechanisms of Zirconia‐Based Toughened Ceramic CrownsInternational Journal of Applied Ceramic Technology · 2026
- Fracture resistance of zirconia fixed dental prostheses after low-temperature degradation and/or cyclic fatigue.Journal of prosthodontics : official journal of the American College of Prosthodontists · 2026
- Fatigue behavior of multilayered zirconias: Impact of finishing and polishing protocols.The Journal of prosthetic dentistry · 2026
- Evaluation of Marginal Fit and Bacterial Adhesion at the Zirconia Crown–Titanium Base Abutment InterfaceClinical Implant Dentistry and Related Research · 2026