Validity of Bone Mineral Density Measurements Using SONIALVISION G4: A Quantitative CT-based Evaluation.
Junichiro Morita, Keisuke Uemura, Sotaro Kono, Yi Gu, Yuto Nishioka, Tomoki Asano, Ryo Higuchi, Hirokazu Mae,
Department of Orthopaedic Surgery, National Hospital Organization Mito Medical Center, Ibarakimachi, Ibaraki, Japan. · Department of Orthopaedic Medical Engineering, Osaka University Graduate School of Medicine, Suita, Osaka, Japan. surmountjp@gmail.com.
Key points
From the paper's conclusion- SONIALVISION G4 showed small differences in femoral neck BMD measurements compared with CT-derived measurements but may systematically overestimate total hip BMD. Further validation studies are warranted.
Abstract
INTRODUCTION: SONIALVISION G4 is a fluoroscopic imaging system equipped with a bone densitometry function calibrated against GE Healthcare dual-energy X-ray absorptiometry (DXA) measurements. However, evidence supporting the validity of in vivo bone mineral density (BMD) measurements remains limited. This study compared BMD measurements obtained using SONIALVISION G4 with GE DXA reference data and evaluated their validity using quantitative computed tomography (CT). MATERIALS AND METHODS: A total of 887 BMD examinations performed with SONIALVISION G4 in women aged ≥45 years were compared with published Japanese reference data obtained using GE DXA. CT-derived BMD was quantified using a validated deep learning-based method and compared with measurements obtained using SONIALVISION G4 in 30 patients. To validate the CT method against GE DXA, 697 patients from four additional institutions underwent GE DXA and CT. Analyses were conducted for both the femoral neck and total hip. RESULTS: Femoral neck BMD generally showed small differences from the reference values, whereas total hip BMD was consistently higher in women aged ≥55 years. CT-derived BMD correlated with SONIALVISION G4 measurements (r = 0.762 and 0.842 for the femoral neck and total hip, respectively), with median absolute differences of 0.022 and 0.133 g/cm2. In contrast, CT-derived BMD showed very strong correlations with GE DXA (r = 0.913 and 0.959) and smaller absolute differences (0.043 and 0.033 g/cm2). CONCLUSION: SONIALVISION G4 showed small differences in femoral neck BMD measurements compared with CT-derived measurements but may systematically overestimate total hip BMD. Further validation studies are warranted.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The association between bone density evaluated on CT or MRI versus the gold standard dual-energy X-ray absorptiometry (DEXA) in the lumbar spine and hip: a meta-analysis.Brain & spine · 2026
- Site-Specific Material Decomposition Versus Universal Z-Effective: Optimizing Opportunistic Bone Mineral Density Screening via Dual-Energy CT.Journal of imaging · 2026
- Site-Specific Associations Between CBCT-Derived Gray Value and Fractal Dimension and DEXA T-Scores in Postmenopausal Women: A Cross-Sectional Analytical Study.Clinical and experimental dental research · 2026
- Age-related bone mineral density patterns in a large German clinical DXA cohort: device-specific comparisons with international reference populations.Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA · 2026
- Combining universal access with high precision for revolutionary osteoporosis screening: CT-based 3D-HU volumetric bone density.Journal of neurosurgery. Spine · 2026
- CT Radiomics and Machine Learning for Three-Class Bone Density Classification: An Exploratory Comparison Across Lumbar Vertebral Levels.Tomography (Ann Arbor, Mich.) · 2026