A molecularly imprinted electrochemical sensor for detection of bisphenol A based on dendritic Pt-Pd bimetallic nanoparticles and g-C3N4 nanocomposite.
Siying Wu, Tianmei Du, Mengyun Ma, Fang Pu, Yanli Zhang, Xue Men, Hongbin Wang, Wenrong Yang,
Functional Nanomaterial-based Chemical and Biological Sensing Technology Innovation Team of Department of Education of Yunnan Province, Yunnan Minzu University, Kunming, 650504, PR China. · Functional Nanomaterial-based Chemical and Biological Sensing Technology Innovation Team of Department of Education of Yunnan Province, Yunnan Minzu University, Kunming, 650504, PR China. ylzhang@ymu.edu.cn.
Key points
From the paper's conclusion- The development of a novel molecularly imprinted electrochemical sensor for the sensitive and selective detection of bisphenol A (BPA) is reported, based on a nanocomposite comprising dendritic Pt-Pd bimetallic nanoparticles and graphitic carbon nitride (Pt-Pd NPs/g-C3N4).
Abstract
The development of a novel molecularly imprinted electrochemical sensor for the sensitive and selective detection of bisphenol A (BPA) is reported, based on a nanocomposite comprising dendritic Pt-Pd bimetallic nanoparticles and graphitic carbon nitride (Pt-Pd NPs/g-C3N4). The molecularly imprinted polymer (MIP) was fabricated via electrochemical polymerization on Pt-Pd NPs/g-C3N4 nanocomposite modified screen-printed electrode (Pt-Pd NPs/g-C3N4/SPE) surface. The synergistic integration of porous Pt-Pd NPs and ultrathin g-C3N4 nanosheets provides a larger effective surface and enhances catalytic activity for BPA determination. The MIP film was fabricated via cyclic voltammetry on the modified electrode, employing BPA as the template molecule and pyrrole as the functional monomer. Following template removal, the resulting molecularly imprinted cavities within the MIP demonstrated high specificity for BPA recognition. The fabricated MIP-based electrochemical sensor exhibited an excellent linear response and high sensitivity for BPA detection over a broad concentration range of 2-100 ng/mL, achieving a limit of detection (LOD) of 0.2 ng/mL (3σ/S) and a limit of quantification (LOQ) of 0.6 ng/mL. The MIP-based sensor exhibits outstanding specificity, reproducibility, and operational stability, thereby providing a versatile platform for rapid and ultrasensitive BPA detection-a critical capability for ensuring environmental safety and protecting public health.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A dual-recognition paper-based sensor integrating a Ru/Co nanoparticle-single-atom catalytic interface for trace bisphenol A monitoring in complex matrices.The Analyst · 2026
- Spectroscopic mechanism-driven design of a magnetic metal-organic framework (M-MXene@Fe-BTC) electrochemical sensor for detection of bisphenol A in water.Analytical methods : advancing methods and applications · 2026
- Fe3O4@NA@MIPs-based molecularly imprinted nanozyme colorimetric platform with smartphone-assisted RGB readout for BPA detection.Analytical methods : advancing methods and applications · 2026
- Differential ratiometric fluorescence sensing enabled by molecularly imprinted dual-emission carbon dots for ultrasensitive bisphenol A determination.The Analyst · 2026
- Sensitive and Selective Electrochemical Detection of Bisphenol A Using Chitosan-Coated Magnetic NanoparticlesJournal of The Electrochemical Society · 2026
- Hybrid CTAB–Silica and APTES-Functionalized Multi-Walled Carbon Nanotubes for Adsorption-Enhanced Electrochemical Detection of Bisphenol A in Environmental and Food SamplesJournal of The Electrochemical Society · 2026