Accuracy of anatomical landmarks for the pelvic C-clamp: a three-dimensional CT imaging study.
Kanyakorn Riewruja, Pannathorn Itdhiamornkulchai, Numphung Numkarunarunrote, Yuda Chongpison, Chavarin Amarase, Sanzhar Artykbay, Saran Tantavisut
Chulalongkorn University and King Chulalongkorn Memorial Hospital, Bangkok, Thailand. · Chulalongkorn University, Bangkok, Thailand.
Key points
From the paper's conclusion- Point A demonstrated the highest observed imaging-defined accuracy and significantly higher odds of both IN and CENTER placement than point C after accounting for patient-level clustering. Point B showed higher odds than point C for both outcomes, but these differences did not reach statistical significance. These findings should be interpreted as bone-referenced, imaging-based estimates rather than direct evidence of biomechanical superiority or clinical safety. Fluoroscopic guidance remains advisable when clinically feasible.
Abstract
PURPOSE: To evaluate the accuracy of three commonly used anatomical landmarks for pelvic C-clamp pin insertion using three-dimensional (3D) CT imaging. METHODS: From a cohort of patients undergoing CT angiography of the lower extremities, 3D bony reconstruction images were created. Three commonly used anatomical insertion landmarks-the intersection points described by Ganz et al. (point A), by Pohlemann et al. (point B), and by the DePuy Synthes Pelvic C-Clamp surgical guide (point C)-were drawn on the 3D CT reconstruction images, and their relationship with the SI joint surface was recorded. We additionally stratified landmark accuracy by SI joint surface morphology. Patient-level clustering was accounted for using generalized estimating equations (GEE). RESULTS: A total of 101 hemipelves were obtained from 51 patients (36 men and 15 women; mean age 61.56 years). In terms of accuracy, point A demonstrated 69.31% of points overlying the SI surface, recorded as 'IN', and 22.77% overlying the central quadrant of the SI surface, recorded as 'CENTER'; point B demonstrated 68.32% and 21.78%; and point C demonstrated 56.44% and 9.90%, respectively. Point A demonstrated significantly higher odds than point C for both 'IN' (OR 1.74, 95% CI 1.03-2.95; p = 0.0377) and 'CENTER' placement (OR 2.69, 95% CI 1.07-6.76; p = 0.0358), whereas the corresponding differences for point B did not reach statistical significance ('IN': OR 1.67, 95% CI 0.99-2.79; p = 0.0534; 'CENTER': OR 2.54, 95% CI 0.91-7.10; p = 0.0761). Anterior misplacement was the most frequent type of malposition for all insertion points. Sacral morphology did not statistically influence the accuracy of any landmark. CONCLUSION: Point A demonstrated the highest observed imaging-defined accuracy and significantly higher odds of both IN and CENTER placement than point C after accounting for patient-level clustering. Point B showed higher odds than point C for both outcomes, but these differences did not reach statistical significance. These findings should be interpreted as bone-referenced, imaging-based estimates rather than direct evidence of biomechanical superiority or clinical safety. Fluoroscopic guidance remains advisable when clinically feasible. LEVEL OF EVIDENCE: III Diagnostic Level III.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- TP4.3 Accuracy of Patient Specific Guide to Achieve Pre-Operatively Templated Correction of Hip Dysplasia with PAOJournal of Hip Preservation Surgery · 2026
- Anatomical fit of a precontoured suprapectineal plate in japanese patients with acetabular fractures: A three-dimensional morphometric study.Injury · 2026
- TP4.5 Precision Without Navigation: Standardized C-Arm Positioning Improves Accuracy in Periacetabular Osteotomy Surgical ModelJournal of Hip Preservation Surgery · 2026
- Three-Dimensional CT-Based Simulation of Four Central Peg Placement Strategies for a Reverse Glenoid Baseplate.Journal of visualized experiments : JoVE · 2026
- EP4.10 Accuracy of Semi-Automated 3D CT Hip Measurement for Femoroacetabular Impingement: Validation Against Surgeon and Engineering Reference StandardsJournal of Hip Preservation Surgery · 2026
- Three-Dimensional Evaluation of Locking Plate Fit to the Distal Humerus.Orthopaedics & traumatology, surgery & research : OTSR · 2026