Automated intracorporeal tool exchange with the Symphera multi-tool prototype: a crossover evaluation of efficiency and surgeon workload in a porcine laparoscopic model.
Fabian Haak, Conrad Heid, Brett Colton Parker, Antonello Forgione, Jacques Marescaux
Division of Hepatobiliary Surgery and Visceral Surgery, Department of Visceral, Transplant, Thoracic and Vascular Surgery, University Hospital Leipzig, Liebigstraße 24, 04103, Leipzig, Germany. Fabian.haak@medizin.uni-leipzig.de. · Symphera GmbH, Munich, Germany.
Key points
From the paper's conclusion- Symphera demonstrated favourable trends in procedural efficiency, gaze behaviour, and subjective workload versus conventional instrument exchange. The convergence of behavioural, subjective, and efficiency data supports the hypothesis that intracorporeal exchange preserves visual attention and may reduce perceived workload. Larger, adequately powered studies are warranted.
Abstract
BACKGROUND: Conventional laparoscopic instrument exchange requires repeated trocar withdrawal, redirecting surgeon gaze and disrupting operative flow. The Symphera system enables intracorporeal tool-tip switching, but its impact on procedural efficiency, visual attention, and workload has not been evaluated in a controlled setting. METHODS: Two surgical fellows performed three standardised laparoscopic tasks (cholecystectomy, controlled hepatic bleeding, dissection-coagulation relay) under Symphera and conventional conditions in a live porcine crossover model. Primary outcomes were total task time and tool change time; secondary outcomes included gaze behaviour (eyes-on proportion, disruption events), subjective workload (SURG-TLX), device usability, and heart rate. RESULTS: Mean total task time was 27.3 ± 10.9 vs. 33.9 ± 13.2 min (-19.6%; HL estimate -6.7 min, 95% CI -12.4 to -1.5; p = 0.10). Tool change time was reduced by 61.5% (3.6 ± 2.3 vs. 9.4 ± 6.6 min; HL estimate -5.2 min, 95% CI -9.7 to -3.4; p = 0.10). Intracorporeal exchanges were completed in a mean of 4.3 ± 4.4 s (75% within 5 s) versus 25.3 ± 23.4 s for extracorporeal exchanges. Eyes-on proportion was higher with Symphera (86.0 ± 10.9% vs. 78.1 ± 9.8%; HL estimate + 7.7 pp, 95% CI + 0.9 to + 14.8; p = 0.031), with instrument exchange as a gaze disruption cause markedly reduced. SURG-TLX scores were 44.8% lower with Symphera (40.2 ± 15.8 vs. 72.9 ± 19.8), with reductions across all six dimensions. CONCLUSION: Symphera demonstrated favourable trends in procedural efficiency, gaze behaviour, and subjective workload versus conventional instrument exchange. The convergence of behavioural, subjective, and efficiency data supports the hypothesis that intracorporeal exchange preserves visual attention and may reduce perceived workload. Larger, adequately powered studies are warranted.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Bedside Assistant Suction Response and Visual Field Recovery During Robot-Assisted Rectal Cancer Surgery: An Exploratory Propensity Score-Matched Comparison of Assistant-Access Workflows Using Hinotori and da Vinci Xi.Asian journal of endoscopic surgery · 2026
- Foot‐Pedal Monitoring Camera Improves Pedal Operation During Simulated Therapeutic EndoscopyDEN Open · 2026
- Adaptive Motion Scaling With Phase‐and‐Gaze Awareness for Laparoscopic Robotic TeleoperationAdvanced Robotics Research · 2026
- An Open-Source da Vinci X/Xi Instrument Module for Laparoscopic Surgery ResearcharXiv · 2026
- Comparative Ergonomic Assessment of Microscopic, Endoscopic, and RobOtol®-Assisted Ear Surgery: A Prospective Pilot Study.Journal of clinical medicine · 2026
- Exploring performance and trust in autonomous surgical robotic systems: the role of visuo-haptic guidance.International journal of computer assisted radiology and surgery · 2026