Langenbeck's archives of surgery

Device-assisted fascial closure of 10-mm or larger trocar sites in laparoscopic gastrointestinal surgery: a prospective non-randomized controlled study.

Jinhu Chen, Shangyi Luo, Jianping Jiang, Yi Zeng, Xiaopeng Wang, Zhengxing Lin, Yueming Lin, Xiaoling Chen,

Department of Gastric Surgery, Clinical Oncology School of Fujian Medical University, Fujian Cancer Hospital, Fuzhou, 350014, China. · Interdisciplinary Institute for Medical Engineering, Fuzhou University, Fuzhou, 350108, China.

Randomised · n=389

Key points

From the paper's conclusion
  • Device-assisted closure was associated with fewer eligible-site technical failures. Non-randomized allocation, sparse events, and potential operator confounding preclude claims of device superiority or hernia prevention. Findings from the concurrent extraction-wound protocol were separate exploratory results and should not be attributed to the closure forceps.

Abstract

PURPOSE: To evaluate the technical reliability of device-assisted fascial closure at eligible non-extraction trocar sites 10 mm or larger. Peri-umbilical extraction-wound outcomes were assessed separately as secondary exploratory outcomes. METHODS: This single-centre prospective non-randomized controlled study included 389 consecutive patients undergoing laparoscopic gastric or colorectal surgery. The primary comparison was device-assisted versus conventional direct-vision fascial closure at eligible non-extraction trocar sites 10 mm or larger (252 vs. 137 patients). A standardized peri-umbilical extraction-wound protocol was concurrently used in the intervention group but was treated as a separate co-intervention. The primary outcome was patient-level fascial-closure failure at an eligible trocar site. Trocar-site hernia was secondary; extraction-wound morbidity and patient-reported cosmetic satisfaction on postoperative day 30 were exploratory. RESULTS: The groups differed in tumour site, extraction-incision location, and incision length. Eligible-site closure failure occurred in 0/252 intervention patients and 4/137 controls (risk difference, -2.9 percentage points; 95% confidence interval, -7.3 to -0.6; P = 0.015). A separate 5-mm drain-tract event occurred in one intervention patient. Trocar-site hernia was uncommon (0 vs. 1 case; P = 0.352). Exploratory extraction-wound hernia and wound-morbidity outcomes did not differ significantly; patient-reported cosmetic satisfaction was higher in the intervention group in an unadjusted comparison (mean difference, 0.30 points; 95% confidence interval, 0.20-0.40; P < 0.001). CONCLUSION: Device-assisted closure was associated with fewer eligible-site technical failures. Non-randomized allocation, sparse events, and potential operator confounding preclude claims of device superiority or hernia prevention. Findings from the concurrent extraction-wound protocol were separate exploratory results and should not be attributed to the closure forceps.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers