Thermal determinants during in situ lithotripsy of nonrelocatable impacted ureteral calculi using a flexible and navigable suction ureteral access sheath: an in vitro study.
Zhaoxin Ying, Litong Zhan, Chaoyue Lu, Yuchen Xie, Na Yu, Yonghan Peng, Shaoxiong Ming, Xiaofeng Gao
Department of Urology, Changhai Hospital, First Affiliated Hospital of Naval Medical University, Shanghai, China. · Department of Urology,, Jinling Hospital, Affiliated Hospital of Medical School, Nanjing University, Nanjing, China.
Key points
From the paper's conclusion- When in situ lithotripsy is unavoidable for nonrelocatable impacted ureteral calculi, laser setting and irrigation flow rate are major determinants of intraluminal temperature elevation. Under the settings shared by both platforms, TFL was associated with greater temperature elevation than Ho: YAG. Adequate irrigation is particularly important at higher laser settings, whereas stronger suction should not be assumed to provide additional thermal protection.
Abstract
PURPOSE: In situ lithotripsy may be unavoidable when impacted ureteral calculi cannot be safely relocated. This in vitro study aimed to identify the key determinants of intraluminal temperature elevation when using a flexible and navigable suction ureteral access sheath (FANS) in this setting. METHODS: A standardized silicone-based in vitro urinary system model simulating impacted ureteral calculi was constructed. Real-time temperature monitoring was performed using an intelligent temperature-sensing flexible ureteroscope. Experiments were conducted using holmium: yttrium-aluminum-garnet (Ho: YAG) and thulium fiber laser (TFL) with varying laser settings, irrigation flow rates (50 and 100 mL/min), negative pressure suction levels (none, - 150 mmHg, - 262 mmHg), and sheath sizes (10F and 12.5F). Temperature changes were continuously recorded during 1-minute laser activation. RESULTS: Irrigation flow rate and laser setting showed significant main effects on intraluminal temperature elevation during Ho: YAG lithotripsy (both P < 0.001), whereas sheath size and suction level showed no significant effects. During TFL lithotripsy, sheath size, suction level, irrigation flow rate, and laser setting were all significantly associated with temperature elevation (all P < 0.01), with irrigation flow rate and laser setting showing the largest effect sizes. Under the laser settings shared by both platforms, TFL produced significantly greater intraluminal temperature elevation than Ho: YAG (P < 0.001). For Ho: YAG, temperature elevation was particularly pronounced at the highest tested setting (1.5 J × 20 Hz), while increasing irrigation provided greater cooling at higher laser settings. During TFL lithotripsy, temperature elevation increased with increasing negative-pressure suction, with the - 262 mmHg condition showing significantly greater temperature elevation than the no-suction condition. CONCLUSION: When in situ lithotripsy is unavoidable for nonrelocatable impacted ureteral calculi, laser setting and irrigation flow rate are major determinants of intraluminal temperature elevation. Under the settings shared by both platforms, TFL was associated with greater temperature elevation than Ho: YAG. Adequate irrigation is particularly important at higher laser settings, whereas stronger suction should not be assumed to provide additional thermal protection.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effect of Conventional Versus Flexible and Navigable Suction Ureteral Access Sheaths on Ureteral Temperature Modulation During High-power Ho:YAG Laser Lithotripsy: An In Vitro EvaluationEuropean Urology Open Science · 2026
- A novel 7.5 single-use flexible ureteroscope with in-built temperature and pressure sensors: Is there a limit to reliability and safety? An in vivo experimental study.World journal of urology · 2026
- Intrarenal pressure during ex-vivo flexible ureteroscopy: a pilot experimental study in cadaveric kidneys.World journal of urology · 2026
- Real-time monitoring with DISS and FANS systems to assure safe and effective stone treatment: Insights from ex-vivo experimental study from EAU endourology.World journal of urology · 2026
- Smart Lithotripsy: Lasers, Energy Sources and Efficiency.Current urology reports · 2026
- An in-vitro comparative study of magneto holmium: YAG and fiber dust thulium fiber lasers-primary effects on ablation, retropulsion, fragmentation, and temperature changes using phantom urologic stones.World journal of urology · 2026