Real-time monitoring with DISS and FANS systems to assure safe and effective stone treatment: Insights from ex-vivo experimental study from EAU endourology.
Hayk Grigoryan, Rifat Ergul, Vineet Gauhar, Arkadya Musayelyan, Armen Muradyan, Armen Avoyan, Evangelos Liatsikos, Panagiotis Kallidonis,
Department of Urology, Yerevan State Medical University after Mkhitar Heratsi, 0087, Yerevan, Armenia. · Department of Urology, İstanbul University, Istanbul, Turkey.
Key points
From the paper's conclusion- Both DISS and FANS effectively maintained IRP within a safe range. Controlled, balanced suction was crucial for safety, while excessive negative pressure or prolonged manual irrigation risked collapse and mucosal injury. The 9.2 Fr DISS + 12/14 Fr FANS configuration demonstrated stable pressure control and efficient performance in this experimental model.
Abstract
PURPOSE: To evaluate the real-time safety profile and intrarenal pressure (IRP) dynamics of two suction-assisted ureteroscopic systems, the Direct In-Scope Suction (DISS) and Flexible and Navigable Suction (FANS) ureteral access sheath, in an ex-vivo model. MATERIALS AND METHODS: Four porcine kidney-ureter units were used, harvested from two female pigs (32-34 kg). A 20G intravenous catheter was inserted into the middle calyx for continuous IRP monitoring via a pressure transducer and monitor. Each kidney underwent four experimental configurations combining 7.5 Fr PU3033AH or 9.2 Fr PU400A DISS scopes with 9.5/11.5 Fr or 12/14 Fr FANS sheaths. Irrigation was performed using a manual hand pump, and suction was applied at 100 mmHg either through the ureteroscope or sheath. Measurements included baseline and maximal IRPs, suction-induced changes, mucosal injury, and time to endoscopic stone-free status (ESFS). RESULTS: Sixteen experiments were completed. Activation of suction consistently reduced IRP across all settings. The The 7.5 Fr DISS + 9.5/11.5 Fr FANS configuration achieved the lowest pressures (4-12 mmHg) but caused pelvicalyceal collapse and six mucosal injuries. In contrast, the 9.2 Fr DISS + 12/14 Fr FANS combination-maintained IRP between 12 and 28 mmHg with efficient fragment evacuation and minimal collapse. CONCLUSION: Both DISS and FANS effectively maintained IRP within a safe range. Controlled, balanced suction was crucial for safety, while excessive negative pressure or prolonged manual irrigation risked collapse and mucosal injury. The 9.2 Fr DISS + 12/14 Fr FANS configuration demonstrated stable pressure control and efficient performance in this experimental model.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Hydrodynamic outcomes of suction-assisted retrograde intrarenal surgery under different irrigation, suction, and ureteroscope conditions: an ex vivo porcine study.Urolithiasis · 2026
- Intrarenal Pressure Monitoring and Control Strategies Using Suction in Flexible Ureteroscopy: An EAU Endourology Systematic Review of Technologies and Clinical Outcomes.Journal of endourology · 2026
- Unchecked Wall Vacuum Might Pose a Risk: An In Vitro Experimental Study on the Effects of Negative Wall Pressure on Suction Ureteral Access Sheaths-A PACE Group Study.Journal of endourology · 2026
- Ex Vivo Porcine Kidney Model for Evaluating Flow-Pressure Dynamics and Stone Retrieval in Suction-Assisted Retrograde Intrarenal Surgery.Journal of visualized experiments : JoVE · 2026
- Intrarenal pressure during ex-vivo flexible ureteroscopy: a pilot experimental study in cadaveric kidneys.World journal of urology · 2026
- Suction-assisted ureteroscopy compared with traditional ureteroscopy for renal stones ≤ 2 cm: a systematic review, Bayesian network meta-analysis and meta-regression.World journal of urology · 2026