Skeletal radiology

Low-dose photon-counting detector CT for assessment of joint space narrowing in peripheral inflammatory arthritis: an accuracy and reliability study.

Tamas Munkacsi, Tamas Purczel, Akos Horvath, Janos Norbert Gyebnar, Pal Maurovich-Horvat, Daniel Sandor Veres, Akos Menyhart, Viktoria Gyergyoi,

Department of Radiology, Medical Imaging Centre, Semmelweis University, 2. Koranyi Sandor Utca, 1083, Budapest, Hungary. · Department of Biophysics and Radiation Biology, Semmelweis University, Budapest, Hungary.

Key points

From the paper's conclusion
  • Low-dose PCD-CT demonstrated excellent inter-reader reliability and close agreement with standard-dose PCD-CT for JSW and JSN assessment. Compared with DR, it provided joint space measurements more consistent with standard-dose PCD-CT at a comparable effective radiation dose.

Abstract

OBJECTIVE: To assess the performance of a low-dose high-resolution photon-counting detector CT (PCD-CT) protocol for evaluating joint space width (JSW) and joint space narrowing (JSN) in peripheral arthritis, including inter-reader reliability and comparison with digital radiography (DR) and digital tomosynthesis (DT), using standard-dose PCD-CT as the reference standard. METHODS: Twenty patients (forty hands) were examined using DR, DT, and low- and standard-dose ultra-high-resolution PCD-CT. JSW was measured, and JSN was subjectively scored in the radioulnar, radiocarpal, intercarpal, carpometacarpal, metacarpal, metacarpophalangeal, and interphalangeal joints. Two readers independently assessed all examinations, and inter-reader reliability was evaluated. DR, DT, and low-dose PCD-CT were compared with standard-dose PCD-CT. Effective doses were calculated from dose-length product and dose-area product values. Statistical analysis included linear mixed-effects models and one-way ANOVA. RESULTS: Inter-reader reliability for JSW measurements was excellent across all modalities (ICC 0.988-0.995; 95% CI 0.987-0.995). Compared with standard-dose PCD-CT, DR significantly overestimated JSW (mean difference + 0.279 mm, p < 0.001) and underestimated JSN severity. No significant differences were observed between low- and standard-dose PCD-CT for JSW (mean difference - 0.016 mm, p = 0.302) or JSN scores (mean difference + 0.019, p = 0.533). DT and standard-dose PCD-CT had higher effective doses than DR and low-dose PCD-CT, with no significant dose difference between DR and low-dose PCD-CT (p = 0.993). CONCLUSION: Low-dose PCD-CT demonstrated excellent inter-reader reliability and close agreement with standard-dose PCD-CT for JSW and JSN assessment. Compared with DR, it provided joint space measurements more consistent with standard-dose PCD-CT at a comparable effective radiation dose.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Low-dose photon-counting detector CT for assessment of joint space narrowing in peripheral inflammatory arthritis: an accuracy and reliability study — Science Skim