Use of the MemoSorb Bioresorbable PFO Device in an 11-Year-Old Diver with Phospholamban Mutation and Familial Hypertrophic Cardiomyopathy.
Sara Cooke, Fouad Kaladji, Andres-Rico Armada, Xiangbin Pan, Wei Li, Marilyn McDougall, Carles Bautista-Rodriguez, Alain Fraisse
Paediatric Cardiology and Cardiac Surgery Services, Royal Brompton & Harefield NHS Hospital, Guys and St Thomas' NHS Foundation Trust, London, UK. · National Center of Cardiovascular Diseases, Beijing, China.
Key points
From the paper's conclusion- We report biodegradable device implantation for primary prevention in an asymptomatic 11-year-old boy with a patent foramen ovale, normal biventricular function and phospholamban mutation, conferring an elevated future risk of hypertrophic cardiomyopathy.
Abstract
We report biodegradable device implantation for primary prevention in an asymptomatic 11-year-old boy with a patent foramen ovale, normal biventricular function and phospholamban mutation, conferring an elevated future risk of hypertrophic cardiomyopathy. He was a keen diver with ambitions to be a marine biologist. Following multidisciplinary discussion, closure of the defect was agreed using a bioresorbable device to optimize future transseptal intervention for potential cardiomyopathy. A 34 mm MemoSorb® PFO Occluder (Lepu medical Technology, Beijing, China) was positioned, complicated by complete atrioventricular block, regressing to first degree after removing the device. A 28 mm device was then successfully implanted. Intermittent 2nd degree atrioventricular block occurred post-procedurally but was successfully treated with steroids, with normal Holter ECG a few weeks later. Bubble test echocardiography was negative after 6 months. The patient was allowed to dive again. The MemoSorb ® biodegradable device can be implanted in selected cases. More studies are needed to confirm safety and efficacy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Feasibility and short-term outcomes of transcatheter closure of moderate-to-large perimembranous ventricular septal defects in children using bioabsorbable occluders: a prospective case series.Frontiers in cardiovascular medicine · 2026
- Fully biodegradable versus nitinol occluder devices for transcatheter patent foramen ovale closure: a systematic review.Global cardiology science & practice · 2026
- Bioresorbable vs Metallic Occluders for Transcatheter Atrial Septal Defect ClosureJAMA · 2025
- Initial Clinical Experience with Percutaneous Endocardial Septal Pulsed Field Ablation in Hypertrophic Obstructive Cardiomyopathy: A First-in-Human Case Series.Journal of clinical medicine · 2026
- Suture-Mediated PFO Closure with the NobleStitch EL System: A Systematic Review and Single-Arm Meta-AnalysisEuropean Heart Journal - Valvular and Structural Heart Disease · 2026
- Pulsed Field Ablation for Atrial Fibrillation: Mechanism, Clinical Evidence, and Emerging Applications.Cureus · 2026