Ultrasensitive miRNA-21 biosensing: a metal-free photoinduced ROMP strategy via 3'-end oxidation.

Yunna Li, Shijin Zhou, Zhidan Tian, Jinming Kong, Xueji Zhang

School of Environmental and Biological Engineering, Nanjing University of Science and Technology, Nanjing, 210094, Jiangsu, China. · Department of Pathology, Nanjing First Hospital, Nanjing Medical University, Nanjing, 210006, Jiangsu, China. tianzhidan@foxmail.com.

Key points

From the paper's conclusion
  • These results demonstrate that target-end-triggered photo-ROMP provides an effective enzyme-free polymer amplification strategy for electrochemical miRNA detection in buffer and diluted serum matrices.

Abstract

MicroRNA-21 (miRNA-21) is frequently overexpressed in various solid tumors, making its reliable quantification important for cancer-related biomarker analysis. Herein, we report an electrochemical sensing strategy that integrates RNA 3'-end oxidation with metal-free photo-induced ring-opening metathesis polymerization (photo-ROMP) for miRNA-21 detection. After hybridization with a thiolated DNA probe immobilized on a gold electrode, the 3'-terminal cis-diol group of miRNA-21 was selectively oxidized by sodium periodate to generate reactive aldehyde groups. The vinyl ether initiator 3-aminopropanol vinyl ether (APVE) was subsequently anchored to the oxidized miRNA-21 through aldehyde-amine condensation followed by NaCNBH3-mediated reductive amination. Under blue-light irradiation, photoexcited 2,4,6-tris(4-methoxyphenyl)pyrylium tetrafluoroborate (MeO-TPT) activated the interfacial APVE and initiated photo-ROMP of the cyclopentadienyl ferrocene (CFc), producing electroactive polymer chains on the electrode surface and amplifying the square-wave voltammetric response. Under optimized conditions, the biosensor exhibited a linear response toward miRNA-21 from 100 aM to 100 pM, with a calculated limit of detection of 28.48 aM. The sensor also showed good sequence selectivity, reproducibility, and short-term stability. Matrix-matched calibration in 10% diluted human serum yielded recoveries of 106.2-114.5% with relative standard deviations of 2.37-4.99%. These results demonstrate that target-end-triggered photo-ROMP provides an effective enzyme-free polymer amplification strategy for electrochemical miRNA detection in buffer and diluted serum matrices.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers