Reliability of Offline Measurements of Stored Transperineal Ultrasound Images: An Exploratory Analysis by Observer Experience.

Ting Xiao, Yunqing Cao, Siyun Du, Zhuoru Luo, Weijun Huang

Ultrasound Diagnosis and Treatment Centre, The First People's Hospital of Foshan (Foshan Hospital Affiliated to Southern University of Science and Technology), Foshan, 528000, China. · Ultrasound Diagnosis and Treatment Centre, The First People's Hospital of Foshan (Foshan Hospital Affiliated to Southern University of Science and Technology), Foshan, 528000, China. hwjun1716@163.com.

Observational · n=398

Key points

From the paper's conclusion
  • Interobserver reliability of offline measurements of stored TPUS images was highest for BP and BND and lowest for RVA. In exploratory analyses, reliability estimates for URA and RVA were lower among junior observers, whereas BP and BND remained highly reliable.

Abstract

INTRODUCTION AND HYPOTHESIS: Previous studies have evaluated the reproducibility of transperineal ultrasound (TPUS) parameters for stress urinary incontinence (SUI); however, their relative reliability remains incompletely characterized. We aimed to evaluate inter- and intra-observer reliability of four TPUS parameters measured offline on stored images and explore differences by observer experience. METHODS: This retrospective study included 398 women who underwent TPUS 6-8 weeks postpartum. Four observers, all radiologists (two senior and two junior), independently assessed four parameters: bladder position (BP), bladder neck descent (BND), urethral rotation angle (URA), and retrovesical angle (RVA) on stored dynamic midsagittal cine loops. Measurements were repeated after 2 weeks using the same image dataset. Reliability was assessed using intraclass correlation coefficients (ICCs), standard error of measurement (SEM), minimal detectable change (MDC95), and Bland-Altman analysis; analyses stratified by observer experience were exploratory. RESULTS: Interobserver ICCs were highest for BP 0.981 (95% CI 0.965-0.988) and BND 0.977 (0.971-0.982), lower for URA 0.850 (0.827-0.871), and lowest for RVA 0.694 (0.653-0.733). SEM/MDC95 were 1.33/3.68 mm for BP, 1.43/3.96 mm for BND, 9.69/26.87° for URA, and 13.69/37.94° for RVA. Stratified by observer experience, ICC estimates for URA and RVA were lower among junior observers, whereas BP and BND remained high in both groups. CONCLUSIONS: Interobserver reliability of offline measurements of stored TPUS images was highest for BP and BND and lowest for RVA. In exploratory analyses, reliability estimates for URA and RVA were lower among junior observers, whereas BP and BND remained highly reliable.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers