Failure of intramedullary nailing in pertrochanteric femur fractures - A retrospective analysis of 383 patients over a 5.5-year period.
Fynn Erik Kankel, Sabine Rademann, Nikola Vukelic, Oliver Pieske
Carl von Ossietzky Universität Oldenburg School VI - School of Medicine and Health Sciences, Ammerländer Heerstraße 114-118, 26129, Oldenburg, Germany. fynn.erik.kankel@uni-oldenburg.de. · Department of Trauma Surgery Orthopaedics and Sports Traumatology, Evangelisches Krankenhaus Oldenburg, Steinweg 13-17, 26122, Oldenburg, Germany.
Key points
From the paper's conclusion- Fracture morphology described by the AO/OTA classification emerged as the strongest predictor of implant failure after intramedullary nailing of pertrochanteric fractures. Unstable subtypes require careful preoperative planning and follow-up. Female sex was also independently associated with implant failure.
Abstract
PURPOSE: Pertrochanteric femoral fractures are a rising health and economic burden in ageing societies. Intramedullary nailing is standard treatment, yet implant failure remains a serious complication. This study evaluated patient, fracture and surgical factors associated with implant failure. METHODS: This single-centre retrospective cohort study included adults who underwent primary intramedullary nailing for AO/OTA 31A1.2, 31A1.3, 31A2.2 and 31A2.3 pertrochanteric fractures at Evangelisches Krankenhaus Oldenburg, a Level I trauma centre in Oldenburg, Germany, between 1 January 2019 and 31 July 2024. Demographic, clinical and radiographic data were extracted from electronic records and imaging archives. Variables included age, sex, documented osteoporosis, AO/OTA classification, reduction quality and tip-apex distance. Implant failure, including cut-out, migration or loosening was the primary endpoint. Associations were tested using Chi-square and multivariable logistic regression. RESULTS: A total of 383 patients (mean age 80.3 years; 70% women) met inclusion criteria. Osteoporosis was documented in 59%. Implant failure occurred in 4.4% (n = 17), pseudarthrosis in 1.0% (n = 4) and revision surgery in 5.5% (n = 21). In the bivariate Chi-square analysis, fracture morphology was the only parameter significantly associated with implant failure (p < 0.001). In multivariable logistic regression, the model was significant (p = 0.012; Nagelkerke R² = 0.202). Female sex (OR 5.82, p = 0.038) and fracture morphology (p = 0.026) were independent predictors. CONCLUSION: Fracture morphology described by the AO/OTA classification emerged as the strongest predictor of implant failure after intramedullary nailing of pertrochanteric fractures. Unstable subtypes require careful preoperative planning and follow-up. Female sex was also independently associated with implant failure. LEVEL OF EVIDENCE: Level III, retrospective cohort study (Oxford 2011). TRIAL REGISTRATION: German Clinical Trials Register (DRKS), DRKS00024627, retrospectively registered on 29 March 2021.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Risk Factors for Failure of Internal Fixation in Postmenopausal Patients With Osteoporotic Intertrochanteric Fractures.Annali italiani di chirurgia · 2026
- Intramedullary nailing compared with primary total hip arthroplasty for pertrochanteric fractures in autonomous elderly patients : a prospective randomized study.Bone & joint open · 2026
- A tale of two nails: A retrospective cohort comparison of TFNA and PFNA in proximal femoral fracture fixation.Injury · 2026
- Peri-Implant Femoral Fractures Around Intramedullary Nails: A Multicenter Case Series.Journal of personalized medicine · 2026
- Cement Augmentation of the Blade in Proximal Femoral Nailing for Trochanteric Fractures in Elderly Patients: A Retrospective Comparison of Mechanical Stability and ComplicationsJournal of Clinical Medicine · 2025
- Catastrophic Failure of a Prosthetic Femoral Stem due to Component Malposition: A Case Report.Case reports in orthopedics · 2026