Clinical Implementation and Performance of Real-Time AI System for Device Tracking in Neurointervention: First U.S. Evaluation of Neuro-Vascular Assist.
João Victor Sanders, Joshua Jimenez, Marion Oliver, Kenichi Kono, Krishna C Joshi, Demetrius Lopes
Brain and Spine Institute, Advocate Health, Chicago, United States. souzasan@musc.edu. · Brain and Spine Institute, Advocate Health, Chicago, United States.
Key points
From the paper's conclusion- Neuro-Vascular Assist was successfully implemented and demonstrated preliminary accuracy and feasibility in a U.S.-based neurointerventional setting. Larger prospective studies are warranted to determine whether real-time AI-assisted device tracking influences procedural decision-making, workflow, safety, or clinical outcomes.
Abstract
PURPOSE: To evaluate the first U.S.-based implementation, notification accuracy, and temporal operator response of Neuro-Vascular Assist. METHODS: In a single-center study, 21 consecutive neurointerventional cases were performed using the AI system integrated with a biplane angiographic system. Procedures were classified as "Standard", "Coil", "Filter", or "Venous". Neuro-Vascular Assist tracked devices including guidewires, stents, coils, and filters in real-time. Visual and audio notifications were made based on device entry and exit in a determined region of interest. We classified these notifications as true positives (TP), false positives (FP), and false negatives (FN), and further assessed for temporal operator response (TOR). RESULTS: The AI system was successfully integrated into the workflow without procedural delays or system failures affecting notification delivery. A total of 232 TPs, 13 FPs, and 17 FNs were recorded, resulting in an overall precision of 94.7% and recall of 93.2%. Among the 59 TP notifications evaluated for TOR, 25 (42.4%) were followed by an observable operator response within 10 s. These 25 positive TOR events represented 10.2% of all 245 notifications. CONCLUSION: Neuro-Vascular Assist was successfully implemented and demonstrated preliminary accuracy and feasibility in a U.S.-based neurointerventional setting. Larger prospective studies are warranted to determine whether real-time AI-assisted device tracking influences procedural decision-making, workflow, safety, or clinical outcomes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Airway management during missions of a neurologist-staffed mobile stroke unit: a retrospective registry study.Scandinavian journal of trauma, resuscitation and emergency medicine · 2026
- Artificial Intelligence for Computer-Aided Detection in Endovascular Interventions: Clinical Applications, Validation, and Translational PerspectivesBioengineering · 2026
- Real-Time Artificial Intelligence-Based Detection of Flow-Directed Microcatheter Tip Markers.Journal of neuroendovascular therapy · 2026
- A Randomized Trial of Telemedicine Models of Care on a Mobile Stroke UnitNEJM Evidence · 2025
- Remote Teleoperation of Endovascular Robots: A Network Emulation Framework for Systematic Performance EvaluationarXiv · 2026
- Feasibility of a novel multi-device robotic system for multiple neurointerventional procedures: pre-clinical validation in porcine models.Journal of robotic surgery · 2026