SEPARATION SCIENCE PLUS

An Equivalency Study Comparing Gas‐assisted Dynamic Accelerated Solvent Extraction With Method 1633A for Analysis of Per‐ and Poly‐Fluorinated Alkyl Substances in Soil and Sediment

Gopal Bera, Harry B. McCarty, German Gomez, Yan Liu

Key points

From the paper's conclusion
  • This study presents a comprehensive equivalency evaluation of a gas‐assisted dynamic accelerated solvent extraction (GA‐dASE) workflow, as an alternative to some of the steps described within Method 1633A for the extraction and analysis of per‐ and polyfluorinated alkyl substances in soils and sediments.

Abstract

This study presents a comprehensive equivalency evaluation of a gas‐assisted dynamic accelerated solvent extraction (GA‐dASE) workflow, as an alternative to some of the steps described within Method 1633A for the extraction and analysis of per‐ and polyfluorinated alkyl substances in soils and sediments. The GA‐dASE workflow automates the extraction step, otherwise labor‐intensive in Method 1633A. Except for the extraction step, both workflows employ identical solvents, cleanup chemistry, and liquid chromatography–tandem mass spectrometry procedures for 40 per‐ and poly‐fluorinated alkyl substances (PFAS). Comparative analysis across Ottawa sand, soils contaminated with PFAS, and standard reference materials demonstrated strong agreement between the two methods. Native and spike recoveries for most target analytes fell within ±30% of each other, and isotope‐dilution internal standard recoveries generally met Method 1633A acceptance criteria. Notable discrepancies were observed for three compounds when extract pH was not adjusted in the range of 6–7 prior to extract cleanup; however, recoveries met the state‐of‐the‐art workflow performance criteria either when pH was adjusted or by using more appropriate labeled reference compounds. Sediment analysis demonstrated that GA‐dASE recoveries closely matched certified reference material values, with perfluorooctanesulfonic acid concentrations agreeing within 97%. Overall, the GA‐dASE workflow provided a streamlined, automated extraction alternative that maintains analytical accuracy and precision while improving productivity and reducing manual labor.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

An Equivalency Study Comparing Gas‐assisted Dynamic Accelerated Solvent Extraction With Method 1633A for Analysis of Per‐ and Poly‐Fluorinated Alkyl Substances in Soil and Sediment — Science Skim