Development and Validation of a Stability‐Indicating Reversed‐Phase High‐Performance Liquid Chromatographic Method for the Simultaneous Determination of Betamethasone Dipropionate, Chlorocresol, and Related Impurities in Betamethasone 0.05% Cream
Grégoire Vincent, Sofiane Fatmi, Mohamed Skiba, Frederic Bounoure, Valentin Lefébure, Nicolas Milon, Malika Lahiani‐Skiba
Key points
From the paper's conclusion- This study describes the development and comprehensive validation of a unified reversed‐phase high‐performance liquid chromatographic method for the simultaneous quantification of betamethasone dipropionate (BDP), chlorocresol, and BDP‐related impurities in a 0.05% cream formulation, following Analytical Quality by Design principles as described in the International Council for Harmonization (ICH)
Abstract
Topical corticosteroid formulations require rigorous analytical control of the active pharmaceutical ingredient, antimicrobial preservatives, and degradation products. Traditional quality control approaches often employ separate methods for each determination, increasing analysis time and solvent consumption. This study describes the development and comprehensive validation of a unified reversed‐phase high‐performance liquid chromatographic method for the simultaneous quantification of betamethasone dipropionate (BDP), chlorocresol, and BDP‐related impurities in a 0.05% cream formulation, following Analytical Quality by Design principles as described in the International Council for Harmonization (ICH) Q14. Chromatographic separation was achieved on an XBridge Shield RP18 column using gradient elution with water–acetonitrile and dual‐wavelength ultraviolet detection (240 nm for BDP and impurities; 280 nm for chlorocresol). Validation was performed according to ICH Q2(R1) guidelines. Baseline resolution (Rs > 2.5) was achieved for all analytes within 50 min. Excellent linearity was demonstrated ( r 2 > 0.999) with y‐intercepts not significantly different from zero. Mean recoveries were 100.1 ± 1.0% for BDP and 100.0 ± 0.9% for chlorocresol. Precision relative standard deviations were below 2.2% for assay and below 4.0% for impurities. Limits of detection and quantification were 0.015% and 0.05%, respectively. Correction factors of 0.83 for betamethasone and 1.09 for beclometasone dipropionate were determined. Nominal operating conditions were selected based on the factorial robustness study as: column temperature 35.0 ± 0.5°C, flow rate 1.50 ± 0.03 mL/min. Solution stability was confirmed for 24 h at 22°C. The proposed unified method replaces three separate analytical procedures with a single chromatographic run, reducing solvent consumption and total analysis time. It is specific, accurate, precise, and robust, offering significant efficiency gains for routine pharmaceutical quality control while maintaining full regulatory compliance.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development and Validation of an Analytical Quality by Design‐Based Robust Liquid Chromatography Method for Determination of Eight Process and Degradation Impurities in Corticosteroid Drug ProductSEPARATION SCIENCE PLUS · 2026
- Green and lean: A validated RP-HPLC-VWD method for simultaneous determination of hydroquinone, methylparaben, fluocinolone acetonide, and propylparaben.Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2026
- A Robust Stability‐Indicating RP‐HPLC Method for Simultaneous Determination of Widely Used Pharmaceutical and Cosmetic PreservativesBiomedical Chromatography · 2026
- Analytical Quality by Design Guided Development of a Sustainable LC–MS Compatible Reverse Phase HPLC Method for Low‐Level Quantification of Lifitegrast and Bromfenac in Ophthalmic Formulations and Simulated Tear FluidSEPARATION SCIENCE PLUS · 2026
- Response Surface Methodology-Assisted Development and Validation of a Stability-Indicating HPLC-PDA Method for Quantification of Medroxyprogesterone Acetate in Microcrystals.Drug design, development and therapy · 2026
- AQbD-Based HPLC Method Development, LC-MS/MS Characterization, and Bivariate Conformity Assessment for Estradiol Valerate and Norethisterone Enanthate in Injectable ContraceptivesJournal of Pharmaceutical Innovation · 2026