NMR in Biomedicine

Improved Precision of High b‐Value In Vivo Cardiac Diffusion Tensor Imaging on a Commercial Ultrahigh Performance Gradient MRI System

Moses Cook, Thomas Garrett, Danielle Kara, Yuchi Liu, Owen Lockwood, Dingheng Mai, Shi Chen, Arwa Younis,

Preclinical

Key points

From the paper's conclusion
  • This study compared cDTI precision at b = 500 and b = 1000 s/mm 2 on a commercially available 3 T MRI system equipped with ultrahigh‐performance gradients.

Abstract

Cardiac diffusion tensor imaging (cDTI) enables non‐invasive assessment of myocardial microstructure, but clinical implementation remains limited by low signal‐to‐noise ratio (SNR) and the prolonged echo times associated with higher diffusion weighting. Ultrahigh‐performance gradients may enable higher b ‐values while maintaining clinically feasible acquisition times. This study compared cDTI precision at b = 500 and b = 1000 s/mm 2 on a commercially available 3 T MRI system equipped with ultrahigh‐performance gradients. Twenty healthy volunteers and six patients with cardiac amyloidosis (CA) underwent cDTI at b = 500 and b = 1000 s/mm 2 using identical acquisition parameters except for diffusion encoding strength and echo time. Mean diffusivity (MD), fractional anisotropy (FA), and helix angle transmurality (HAT) were calculated. SNR, geometric distortion, and measurement uncertainty were also assessed using bootstrap analysis. Paired statistical comparisons were performed between diffusion weightings in healthy volunteers. In healthy volunteers, b = 1000 s/mm 2 produced significantly lower MD (1.42 ± 0.05 vs. 1.51 ± 0.05 × 10 −3 mm 2 /s, p < 0.001) and FA (0.30 ± 0.02 vs. 0.32 ± 0.01, p < 0.001) and greater HAT magnitude (−0.86 ± 0.14 vs. −0.75° ± 0.18°/% transmural depth, p < 0.01) compared with b = 500 s/mm 2 . Spatial variance was significantly reduced at b = 1000 s/mm 2 for MD, FA, and HAT. Uncertainty was reduced for MD, FA, and the first eigenvector. In CA, MD was higher and FA and HAT magnitude were lower than in healthy volunteers at both b ‐values. In sum, b = 1000 s/mm 2 cDTI demonstrated reduced spatial heterogeneity and lower measurement uncertainty compared with b = 500 s/mm 2 despite reduced SNR. Higher diffusion weighting may therefore improve the precision and microstructural specificity of cardiac diffusion measurements.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Improved Precision of High b‐Value In Vivo Cardiac Diffusion Tensor Imaging on a Commercial Ultrahigh Performance Gradient MRI System — Science Skim