Magnetic resonance in medicine

High-Resolution Multi-Delay Renal Arterial Spin Labeling Using Radial Imaging at 5T MRI.

Xinyu Tong, Zihan Ning, Ne Yang, Chen Chen, Rui Shen, Shuheng Zhang, Shuo Chen, Zuo-Xiang He,

School of Biomedical Engineering, Tsinghua University, Beijing, China. · Shanghai United Imaging Healthcare, Shanghai, China.

Key points

From the paper's conclusion
  • Radial-ASL at 5T enables free-breathing multi-delay renal perfusion quantification with high spatial resolution and demonstrates good agreement with SAMURAI technique as well as good repeatability.

Abstract

PURPOSE: To develop a high-resolution multi-delay renal arterial spin labeling (ASL) technique for free-breathing imaging at 5T. METHODS: A radial renal ASL (Radial-ASL) sequence was proposed using flow-sensitive alternating inversion recovery (FAIR) preparation followed by continuous golden-angle radial acquisition. The k-space data were firstly sorted by delay time and then reordered based on respiratory motion, with data grouped into four respiratory phases for Spatio-Temporal eXtra-Dimensional Golden-angle RAdial Sparse Parallel (ST-XD-GRASP) reconstruction. Both the Buxton kinetic model and a stepwise kinetic model were applied to perfusion quantification. The proposed method was evaluated against the saturated multi-delay ASL (SAMURAI) technique using Wilcoxon signed-rank tests, intraclass correlation coefficients (ICC), and correlation analysis. Scan-rescan repeatability and interobserver reproducibility were assessed using ICC. RESULTS: Twelve healthy volunteers (mean age: 27.4 ± 4.9 years old, 7 males) and one patient (55-year-old female) were included. Radial-ASL with ST-XD-GRASP provided high-quality images with 1.17 mm in-plane resolution. The stepwise model showed improved fitting performance (R2 ≥ 0.95) compared with the Buxton model (R2: 0.36-0.85). Cortical renal blood flow showed no significant difference between Radial-ASL and SAMURAI (363.7 ± 30.3 vs. 360.4 ± 30.0 mL/100 g/min, p = 0.216), with good agreement (ICC = 0.910) and strong correlation (r = 0.912). Radial-ASL demonstrated good scan-rescan repeatability (ICC = 0.884-0.911) and interobserver reproducibility (ICC = 0.943-0.985). CONCLUSION: Radial-ASL at 5T enables free-breathing multi-delay renal perfusion quantification with high spatial resolution and demonstrates good agreement with SAMURAI technique as well as good repeatability.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

High-Resolution Multi-Delay Renal Arterial Spin Labeling Using Radial Imaging at 5T MRI — Science Skim