Magnetic resonance in medicine

Importance of Background Suppression and Post Label Delay Optimization for White Matter Perfusion Imaging Using Arterial Spin Labeling at 7 Tesla.

Emiel C A Roefs, Icaro A F Oliveira, Suzanne L Franklin, Natalia Petridou, Matthias J P van Osch, Lydiane Hirschler

C.J. Gorter MRI Center, Department of Radiology, Leiden University Medical Center, Leiden, the Netherlands. · Spinoza Centre for Neuroimaging, Netherlands Academy for Arts and Sciences, Amsterdam, the Netherlands.

Key points

From the paper's conclusion
  • BGS becomes increasingly important for perfusion imaging at higher field strengths. Background suppressed WM perfusion imaging at 7 Tesla might benefit from a longer PLD than typically recommended for GM.

Abstract

PURPOSE: Ultra-high field MRI (7 Tesla) could potentially improve white matter (WM) perfusion imaging because of less severe loss of label thanks to longer T1-values. However, physiological noise increases, making background suppression (BGS) likely more important at this field strength. The aim of this study is first to assess the importance of background suppression at 7 Tesla compared to 3 Tesla, and secondly to use that knowledge to optimize flow-alternating-inversion-recovery (FAIR) ASL for WM perfusion imaging. METHODS: A background suppressed FAIR ASL labeling scheme was implemented at both 7 and 3 Tesla scanners. Perfusion temporal SNR was compared over a wide range of BGS-levels by exploiting a 2D multi-slice readout. The knowledge gained from this was used to optimize a FAIR ASL scan with 3D readout scanned at several post label delays (PLD). The tSNR was compared across PLD for GM and several WM masks. RESULTS: BGS was found to be more important at 7 Tesla compared to 3 Tesla. Therefore, a 3D FAIR ASL scan with homogeneous high BGS was used for WM perfusion imaging. Across PLD, we found a parabolic trend for significant voxels in the deep WM, both for a non-zero but slightly stronger for the ischemic perfusion threshold, peaking around a PLD of 2300 ms. However, this was not significantly higher than a PLD of 1800 and 2800 ms in this healthy population. CONCLUSION: BGS becomes increasingly important for perfusion imaging at higher field strengths. Background suppressed WM perfusion imaging at 7 Tesla might benefit from a longer PLD than typically recommended for GM.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Importance of Background Suppression and Post Label Delay Optimization for White Matter Perfusion Imaging Using Arterial Spin Labeling at 7 Tesla — Science Skim