Importance of Background Suppression and Post Label Delay Optimization for White Matter Perfusion Imaging Using Arterial Spin Labeling at 7 Tesla.
Emiel C A Roefs, Icaro A F Oliveira, Suzanne L Franklin, Natalia Petridou, Matthias J P van Osch, Lydiane Hirschler
C.J. Gorter MRI Center, Department of Radiology, Leiden University Medical Center, Leiden, the Netherlands. · Spinoza Centre for Neuroimaging, Netherlands Academy for Arts and Sciences, Amsterdam, the Netherlands.
Key points
From the paper's conclusion- BGS becomes increasingly important for perfusion imaging at higher field strengths. Background suppressed WM perfusion imaging at 7 Tesla might benefit from a longer PLD than typically recommended for GM.
Abstract
PURPOSE: Ultra-high field MRI (7 Tesla) could potentially improve white matter (WM) perfusion imaging because of less severe loss of label thanks to longer T1-values. However, physiological noise increases, making background suppression (BGS) likely more important at this field strength. The aim of this study is first to assess the importance of background suppression at 7 Tesla compared to 3 Tesla, and secondly to use that knowledge to optimize flow-alternating-inversion-recovery (FAIR) ASL for WM perfusion imaging. METHODS: A background suppressed FAIR ASL labeling scheme was implemented at both 7 and 3 Tesla scanners. Perfusion temporal SNR was compared over a wide range of BGS-levels by exploiting a 2D multi-slice readout. The knowledge gained from this was used to optimize a FAIR ASL scan with 3D readout scanned at several post label delays (PLD). The tSNR was compared across PLD for GM and several WM masks. RESULTS: BGS was found to be more important at 7 Tesla compared to 3 Tesla. Therefore, a 3D FAIR ASL scan with homogeneous high BGS was used for WM perfusion imaging. Across PLD, we found a parabolic trend for significant voxels in the deep WM, both for a non-zero but slightly stronger for the ischemic perfusion threshold, peaking around a PLD of 2300 ms. However, this was not significantly higher than a PLD of 1800 and 2800 ms in this healthy population. CONCLUSION: BGS becomes increasingly important for perfusion imaging at higher field strengths. Background suppressed WM perfusion imaging at 7 Tesla might benefit from a longer PLD than typically recommended for GM.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Temporal evolution acquisition based arterial spin labeling (TEA-ASL) for accurate estimation of apparent T2 of perfusion signal in brain.Imaging neuroscience (Cambridge, Mass.) · 2026
- Early post-contrast signal changes in CSF spaces and CSF outflow-related regions on 7 Tesla MRIScientific Reports · 2026
- Hyperintense Signals in Cerebral Blood Flow Maps of Mice Acquired with Pseudo-continuous Arterial Spin Labeling MR Imaging.Magnetic resonance in medical sciences : MRMS : an official journal of Japan Society of Magnetic Resonance in Medicine · 2026
- Motion-robust high-resolution 3D-GRASE pseudo-continuous ASL perfusion imaging at 3TmedRxiv · 2026
- Non‐Contrast Assessment of Blood–Brain Barrier Permeability to Water: Improved Signal Modeling and Data AcquisitionMagnetic Resonance in Medicine · 2025
- Improving Multi-Delay-ASL through specialized reconstruction.ArXiv · 2026