Revisiting the humidity correction - A challenge to the current consensus.
Yuichiro Morishita, Hideki Yamaoka
National Metrology Institute of Japan, AIST, Tsukuba, Ibaraki, Japan. · TOYO MEDIC Co., Ltd., Nishi-ku, Osaka, Japan.
Key points
From the paper's conclusion- Measurements using modern instrumentation indicate that the humidity dependence of k h and the corresponding relative W -value for Co-60 beam quality differ slightly from the long-standing consensus established in ICRU Report 31. Although these differences are of minor importance for practical dosimetry, they may be relevant to primary air-kerma standards. The present results suggest that primary standard laboratories may re-evaluate their uncertainty budgets in light of these findings.
Abstract
BACKGROUND: The humidity correction factor, k h , for ionization chambers in radiation dosimetry has relied on a consensus value established in ICRU Report 31 nearly half a century ago. PURPOSE: This study aims to rigorously re-evaluate this long-standing consensus by employing modern instrumentation. METHODS: First, the equilibration time for the cavity air was examined by implanting a miniaturized humidity sensor directly into the cavity of a broken commercial ionization chamber to address uncertainties regarding stabilization in recent reports. Second, the k h was measured using a vacuum-assisted gas exchange method, in which the air inside the cavity was exchanged actively. Finally, Monte Carlo simulations were performed to analyze energy deposition within the cavity, enabling the conversion of the measured k h to the relative W -value. RESULTS: The direct monitoring of humidity inside the cavity revealed that the equilibration time scale extended to several days, which was much longer than the time scales assumed in the recent reports. The measured k h for the Co-60 beam exhibited slight but apparent differences from the long-standing consensus: the measured k h extrapolated to 0 % relative humidity (RH) was 0.9962-0.9970, and the sudden response increase started from 0 % RH to 40 % RH (or possibly 30 % RH). The derived relative W -value showed a roughly linear dependence on RH, which contrasted with the non-linear behavior of the current consensus curve. CONCLUSIONS: Measurements using modern instrumentation indicate that the humidity dependence of k h and the corresponding relative W -value for Co-60 beam quality differ slightly from the long-standing consensus established in ICRU Report 31. Although these differences are of minor importance for practical dosimetry, they may be relevant to primary air-kerma standards. The present results suggest that primary standard laboratories may re-evaluate their uncertainty budgets in light of these findings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Dose-rate metrics for ionization-chamber recombination in synchrotron-based ultra-high-dose-rate proton and carbon-ion beams: application and validity of quasi-static response models.Zeitschrift fur medizinische Physik · 2026
- Performance of air kerma - area product meters based on long-term calibration data.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Evaluation of the Exradin A30 Parallel Plate Ion Chamber as a Reference Dosimeter in Ultra-High Dose Rate (UHDR) Electron Beams.ArXiv · 2026
- Impact of a common user-derived electrometer calibration coefficient on absorbed dose-to-water determination: a multicenter uncertainty studyJournal of Radioanalytical and Nuclear Chemistry · 2026
- Humidity effects in GafChromicTMEBT4 films for postal audits: characterization, reversibility and exploration of correction methods.Physics in medicine and biology · 2026
- Advancing understanding of temporal performance and behaviour in consumer-grade electronic radon monitors.Radiation protection dosimetry · 2026