Accuracy of Navigated Osteotomies in Rhinoplasty: A Cadaveric Analysis.
Tom A Lauwers, Joel J Vavrina, Christian Thüring, Giovanni Colacicco, Thomas Gander, Kris S Moe, Michael B Soyka
Department of Otorhinolaryngology Head and Neck Surgery, University Hospital Zürich, University of Zürich, Zürich, Switzerland. · Institute of Anatomy, University of Zurich, Zürich, Switzerland.
Key points
From the paper's conclusion- Overall, on imaging, the navigation‐guided osteotomies deviated from the suture by an average of 1.50 mm (95% CI 0.87–2.12 mm) with a standard deviation (SD) of 1.35 mm.
Abstract
BACKGROUND: Navigation guided rhinoplasty allows detailed planning and guidance of osteotomies. We aimed to investigate whether the use of a CT navigation system is feasible and capable of performing more precise osteotomies at specific locations, thus improving accuracy. METHODS: Bilateral electromagnetic-navigated osteotomies (n = 18) of the nasomaxillary suture were performed on nine cadaver heads, using a transnasal (n = 4) or percutaneous (n = 14) technique. After dissection, we manually assessed the distance between the osteotomy and the suture on (n = 5) specimens. Additionally, at baseline and postoperatively, a cone beam computed tomography was performed to calculate the mean and median distance between the osteotomy and the suture line. RESULTS: The manually measured mean distance from intended to actual osteotomy was 0.75 mm (mm) (95% confidence interval [CI] 0.06-1.44 mm) with a median distance of 0.00 mm (interquartile range [IQR] 0.00-1.19 mm). In 6 of 10 procedures, the osteotomy was created according to plan with no visible deviation. Overall, on imaging, the navigation-guided osteotomies deviated from the suture by an average of 1.50 mm (95% CI 0.87-2.12 mm) with a standard deviation (SD) of 1.35 mm. The median distance was 0.73 mm (IQR 0.59-2.61 mm). The accuracies of transnasal and percutaneous osteotomies were similar (p = 0.63). CONCLUSION: Navigation-guided osteotomies demonstrated good feasibility with clinically useful sub-millimeter accuracy for both percutaneous and transnasal techniques. A model for postoperative outcome analysis is described with clinically useful accuracy, though not as high as direct observation. LEVEL OF EVIDENCE: N/A.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effectiveness of Virtual Surgical Planning (VSP) in Craniofacial Surgery : A Systematic Review of Randomized Controlled TrialsThe International Journal of Medical Science and Health Research · 2026
- Augmented reality versus three-dimensionally printed guides in Le Fort I osteotomy: a preclinical study.International journal of oral and maxillofacial surgery · 2026
- Three-Dimensional Accuracy of Static Guides and Dynamic Navigation for Palatal Miniscrew Insertion: A Systematic Review of Template Fabrication and Clinical WorkflowsBioengineering · 2026
- Augmented Reality Navigation Versus Patient-Specific Cutting Guides for Simulated Fronto-Orbital Surface Trajectories in Facial Feminization Surgery: A Preclinical Accuracy Study.Bioengineering (Basel, Switzerland) · 2026
- Surgical cutting and repositioning guides for minimally invasive orthognathic surgery: a retrospective comparative cohort study of bilateral sagittal split osteotomy.International journal of oral and maxillofacial surgery · 2026
- Accuracy of β-tricalcium phosphate gap patient-specific implants in chin osteotomies: a pilot validation study.International journal of oral and maxillofacial surgery · 2026