Factors Associated With Magnet Retention Issues in Cochlear Implants: A Single Institution Analysis.
Sean Haimowitz, Morgan Yun, Anthony Saad, Yu-Lan Mary Ying, Nicole Raia
Department of Otolaryngology, Rutgers New Jersey Medical School, Newark, New Jersey, USA. · Department of Audiology, University Hospital, Newark, New Jersey, USA.
Key points
From the paper's conclusion- Age < 45 and BMI ≥ 30 are useful predictors of high magnet strength or retention issues. Clinicians may use these parameters during preoperative counseling and consider using preoperative CT imaging to assess scalp thickness for these high-risk patients. Physician counseling for these patients should include the likelihood of increased retention issues, especially with OTE processors.
Abstract
OBJECTIVES: To identify clinical and anatomic factors associated with magnet retention issues and high magnet strength requirements in patients undergoing cochlear implantation. METHODS: Retrospective chart review of adult patients (n = 168) who underwent cochlear implantation at a tertiary academic medical center between 2018 and 2023. Data was collected on the following variables: age, BMI at time of surgery, preoperative scalp thickness, magnet type (axial vs. diametric), processor type (behind-the-ear [BTE] vs. off-the-ear [OTE]), and postoperative retention issues (inability to use intended processor, need for compression/hair shaving, delayed activation, or revision surgery). Statistical analysis was performed via chi-square, Fisher's exact, and logistic regression. RESULTS: Out of 168 ears, 13.5% had magnet retention issues. Forty-eight percent of BTE users needed a high magnet strength (> 3) or had retention issues. OTE processors had higher retention issues rates than BTE, especially with diametric magnets. Magnet retention was significantly higher for patients younger than 45 (OR 6.88) and with BMI ≥ 30 (OR 9.85). Those with diametric magnets were at much higher risk than those with axial magnets (17.6% vs. 2.0%) and OTE processors were strongly associated with retention failure (23.1% vs. 4.3%). CONCLUSION: Age < 45 and BMI ≥ 30 are useful predictors of high magnet strength or retention issues. Clinicians may use these parameters during preoperative counseling and consider using preoperative CT imaging to assess scalp thickness for these high-risk patients. Physician counseling for these patients should include the likelihood of increased retention issues, especially with OTE processors. LEVEL OF EVIDENCE: III.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Mid-field MRI with cochlear implants: a healthy volunteer study of magnet displacement, pressure perception, and temporal bone imagingBMC Medical Imaging · 2026
- Key pre-operative CT parameters for cochlear implantation: a focused radiologic review with surgical relevanceNeuroradiology · 2026
- Scalar position characteristics and translocation rates of different lateral wall electrodes within the cochlear basal turn: a cone beam computed tomography study.European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2026
- When They Stop Coming: Key Factors in Pediatric BCHD Adherence in Microtia Patients.The Laryngoscope · 2026
- Quantifying Soft‐Surgery in Cochlear Implantation: Multimodal Data From 30 International SpecialistsThe Laryngoscope · 2026
- Association of Superior Semicircular Canal Roof Thickness With Age and Sex on Photon-counting Detector CT in a Japanese Cohort.Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2026