Scalar position characteristics and translocation rates of different lateral wall electrodes within the cochlear basal turn: a cone beam computed tomography study.
Irumee Pai, Paolo Rondi, Isobel Morley, Emily Hocknell, Andrew Hoey, Sebastien Ourselin, Christos Bergeles, Stephen Connor,
St. Thomas' Hearing Implant Centre, Guy's and St. Thomas' NHS Foundation Trust, 2nd Floor Lambeth Wing, London, UK. irumee.i.pai@kcl.ac.uk. · Department of Radiology, Radiological Sciences and Public Health, University of Brescia, Brescia, Italy.
Key points
From the paper's conclusion- Significant differences in scalar position and STL rates were observed among LW electrodes, with MFX showing the most stable ST placement.
Abstract
PURPOSE: This retrospective study systematically compared scalar position characteristics and translocation (STL) rates within the cochlear basal turn (CBT) across three different lateral wall (LW) electrodes. METHODS: The study included 148 morphologically normal cochleae implanted with Advanced Bionics SlimJ (ABSJ, n = 51), Cochlear Slim Straight (CSS, n = 51), or MED-EL FLEX (MFX, n = 46). Cochlear dimensions (distances A and B) and angular insertion depth (AID) were measured by two observers. Scalar position was assessed on postoperative cone-beam CT at 90°, 180°, 270°, and 360° by three observers using a five-point scale (1-2 = scala vestibuli (SV), 3 = central, 4-5 = scala tympani (ST)). RESULTS: Distances A and B were comparable across groups (p > .05). Mean AID was significantly greater for MFX (p < .001). MFX had higher median scalar scores than CSS at every angle (all Bonferroni-adjusted p < .001) and than ABSJ at 180°, 270°, and 360° (adjusted p < .001). ST retention (all four median scores ≥ 3) was observed in 92.2% of ABSJ, 78.4% of CSS, and 100% of MFX cases, while STL (median score decreased to 1-2 and remained ≤ 3) occurred in 7.8%, 15.7%, and 0%, respectively. MFX demonstrated significantly lower STL and higher ST retention rates than CSS (adjusted p = .018 and p < .003, respectively). No independent associations of distance A, distance B, or AID with either STL or ST retention were identified (all p ≥ .54). CONCLUSION: Significant differences in scalar position and STL rates were observed among LW electrodes, with MFX showing the most stable ST placement.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Partial insertion and electrode array migration in 918 lateral wall cochlear implant electrode arrays: dependence on electrode array design and learning curvesFrontiers in Neurology · 2026
- Comparative Long-Term Outcomes of Lateral Wall and Slim Modiolar Cochlear Implant Electrodes in Children With Normal Cochlear Anatomy.Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2026
- Three-dimensional CT evaluation of anatomical constraints during posterior lip drilling in the retrosigmoid approach to the internal auditory canal.Surgical and radiologic anatomy : SRA · 2026
- The angular trajectory of the vestibular aqueduct: evaluation using photon-counting CT.Insights into imaging · 2026
- Comparison of cone beam computed tomography multiplanar reconstruction in the plane of stapes and the lateral semicircular canal for the diagnosis of otosclerosis.European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2026
- Symons and Fanning Classification in Otosclerosis and its Relation to Audiologic Parameters, Stapes Footplate Thickness, and Round Window Involvement.Audiology & neuro-otology · 2026