Microbubble Timing and Vascular Geometry Are Associated With Vessel-Level Blood-Brain Barrier Opening During Focused Ultrasound.
Seongdong Park, Hyemin Lee, Seunghwan Choi, Juyoung Park, Sun Kwang Kim, Kisoo Kim
Department of Biomedical Engineering, Kyung Hee University, Seoul, Republic of Korea. · Department of Science in Korean Medicine, Graduate School, Kyung Hee University, Seoul, Republic of Korea.
Key points
From the paper's conclusion- Microbubble handling and injection-to-sonication timing are potential protocol variables in FUS-BBBO studies. Control-normalized extravascular metrics provide a reproducible framework for vessel-level longitudinal 2-photon quantification.
Abstract
OBJECTIVES: To examine whether minute-scale microbubble handling and injection-to-sonication timing are associated with vessel-level focused ultrasound (FUS)-mediated blood-brain barrier opening (BBBO), and whether control normalization improves longitudinal 2-photon (2P) measurements. METHODS: Cranial-windowed mice received FITC-dextran (40 kDa) and SonoVue microbubbles by retro-orbital injection, followed by 1.0 MHz FUS at 0.2-0.4 MPa. Two-photon stacks were acquired over 0-15 minutes. Frangi vesselness and Otsu thresholding defined vessel masks and a perivascular ring for extravasation analysis. Three timing conditions were compared: standard (injection within ~5 seconds of reconstitution, sonication immediately); Microbubble Floating (microbubbles held 2 minutes before injection); and Sonication Delay (sonication started 2 minutes after injection). Per-mouse AUC used 2-way ANOVA; slice-level vessel-area dependence used a linear mixed-effects model with mouse as a random intercept. RESULTS: Control intravascular fluorescence decreased by ~24% over 15 minutes, showing that the intravascular signal is not a stable denominator; control-normalized extravascular AUC was therefore the primary metric. Under the standard protocol, extravascular AUC peaked at 0.3 MPa, but the across-pressure difference was not significant (F(2,6) = 3.36, p = .105). Timing effects were descriptive: relative to standard, both Microbubble Floating and Sonication Delay increased extravascular AUC at 0.2 MPa but decreased it at 0.3 MPa. Vessel area was inversely associated with extravasation at the slice level (β = -0.37 per μm2, p = .010), with mouse-level intraclass correlation (ICC = 0.80) indicating dominant between-mouse variability. CONCLUSIONS: Microbubble handling and injection-to-sonication timing are potential protocol variables in FUS-BBBO studies. Control-normalized extravascular metrics provide a reproducible framework for vessel-level longitudinal 2-photon quantification.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Precise Manipulation of Bubble Cavitation and Blood-Brain Barrier Modulation using Temporal Interference Ultrasound.BME frontiers · 2026
- Transcranial microbubble-enhanced ultrafast Doppler imaging of ultrasound-induced brain stimulation using a dual-mode concave low-frequency array transducer.Ultrasonics · 2026
- Imaging strategies to assess efficacy and safety of ultrasound-induced blood-brain barrier openingnpj Acoustics · 2025
- Safe focused ultrasound-mediated blood-brain barrier opening is driven primarily by transient reorganization of tight junctionsCommunications Engineering · 2026
- Early micro- and nanoscale microglial responses to blood-brain barrier modulation by transcranial focused ultrasound.Theranostics · 2026
- Towards Quantitative In Vitro Modelling of Focused Ultrasound-mediated Blood-Brain Barrier Opening using an Ultrasound-transparent Organ-on-Chip DevicebioRxiv · 2026