Safety of bioelectrical impedance analysis in patients with cardiac implantable electronic devices: A systematic review.
H Zweers, Z Zhao, E Noom, W Visser, I Jans, L Willems, M Verhoeven, D A Theuns,
Department of Gastroenterology and Hepatology, Dietetics, Radboud University Medical Center, Nijmegen, The Netherlands. · Department of Internal Medicine, Division of Dietetics, Erasmus MC, Rotterdam, The Netherlands.
Key points
From the paper's conclusion- No evidence of electromagnetic interference or adverse effects on CIED function was identified in the available studies evaluating BIA in patients with CIEDs. These findings support reconsideration of clinical recommendations that discourage BIA use in this population.
Abstract
BACKGROUND: Bioelectrical impedance analysis is widely used in clinical settings to assess body composition, contributes to the management of various health conditions. However, for patients with cardiac implantable electronic devices, such as pacemakers and implantable cardioverter-defibrillators, concerns about electromagnetic interference have led to longstanding recommendations against using bioelectrical impedance analysis. In this systematic review, we aimed to assess the available evidence regarding the safety of cardiac implantable electronic devices (CIEDs) during bioelectrical impedance analysis procedures. METHODS: Following PRISMA guidelines, we searched PubMed, IEEE Xplore, Embase, and CINAHL for studies reporting on bioelectrical impedance analysis safety in patients with cardiac implantable electronic devices. Inclusion criteria focused on human studies reporting CIED function outcomes during BIA measurements. Data on BIA frequencies, types of CIEDs and BIA, and observed electromagnetic interference effects were extracted. The quality and bias of the studies were assessed using the Oxford Levels of Evidence and ROBINS-I tools. RESULTS: Out of 2003 records screened, 8 studies were included, encompassing 866 BIA measurements across frequencies ranging from 5 to 500 kHz. No studies reported electromagnetic interference or adverse events affecting CIED functionality. CONCLUSION: No evidence of electromagnetic interference or adverse effects on CIED function was identified in the available studies evaluating BIA in patients with CIEDs. These findings support reconsideration of clinical recommendations that discourage BIA use in this population.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Safety of Bioelectric Impedance Analysis Testing in Pediatric Patients With Cardiac Implantable Electronic Devices.Pacing and clinical electrophysiology : PACE · 2026
- Bioelectrical Impedance in Cardiovascular Health: Technologies, Clinical Applications, and Future Directions.JACC. Basic to translational science · 2026
- Long-term safety and efficacy of arrhythmia monitoring with an implantable cardiac monitor.International journal of cardiology · 2026
- Human bioimpedance-based state detection technologies for sports health monitoring: a review.Frontiers in physiology · 2026
- Pulsed field ablation of atrial arrhythmias in patients with cardiac implantable electronic devices: a multicenter evaluation of electrical integrity and device safety.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026
- Characteristics, practice patterns of removal, and relapse bacteraemia in patients with a cardiac implantable electronic device (CIED) and bacteraemia: a nationwide studyEuropean Heart Journal · 2026