Measuring Interference, Internal Bone Displacements and Strains Following Cementless Tibial Tray Implantation Using Micro-CT and DVC.
Kieran J Bennett, Lauren S Wearne, Sophie Rapagna, Sunil P Reddy, Julius C Dohm, Christoph Schilling, Berna Richter, Thomas M Grupp,
Medical Device Research Institute, College of Science and Engineering, Flinders University, Adelaide, South Australia, Australia. · The Joint Replacement Clinic, Adelaide, South Australia, Australia.
Key points
From the paper's conclusion- By comparing the results from titanium knee implants with three different designs, we developed a method which can be used for identifying the effect of different implant designs on internal bone mechanics in experimental studies on cadaveric bone following implantation.
Abstract
The primary stability of proximal tibia cementless total knee arthroplasty components is achieved through a bone-implant interference fit. This study aimed to develop the methods to explore this interference fit, and to quantify the resultant internal bone displacement and strain distributions due to implantation with cementless tibial components into human cadaveric proximal tibiae for different implant designs, featuring wings, keel and two types of pegs. Three in-development titanium implants were implanted into three human cadaveric tibiae. Micro-CT images were taken both after resection and after implantation, and image volumes were rigidly registered in the tibial metaphysis. Interference fit was quantified as the distance between the implant boundary in the implanted images and the bone in the resected images. Digital volume correlation was then used to calculate the internal bone displacements and strains. More than 75% of the surface of the implant wing geometries had bone-implant interference (> 0 μm). The total bone displacement due to implantation ranged from 198 to 259 μm and was dominated by bone displacements in the posterior and inferior directions. The 95th percentile minimum principal strain within 5.5 mm of the implant surface ranged from -12,353 to -20,882 μϵ, with 12% to 20% of this volume demonstrating strains which indicate yielding. By comparing the results from titanium knee implants with three different designs, we developed a method which can be used for identifying the effect of different implant designs on internal bone mechanics in experimental studies on cadaveric bone following implantation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Inducible displacement measured with radiostereometric analysis in primary knee arthroplasty: a systematic review of clinical studies.Acta orthopaedica · 2026
- A new cementless tibial baseplate in kinematically aligned total knee arthroplasty has low migration and stabilizes at 3 months: A cohort study using radiostereometric analysis.Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA · 2026
- Native varus angle does not affect the primary fixation of cementless TKA using kinematic alignment - A population-based FE study.Journal of biomechanics · 2026
- Tibial migration of a novel cementless total knee arthroplasty : a multicentre radiostereometric analysis study.Bone & joint open · 2026
- Comparison of CT-based migration analysis against radiostereometric analysis for assessing two-year migration in an uncemented medial pivot tibial knee implant : clinical implications from a prospective safety study.Bone & joint open · 2026
- Tibial implant migration of cemented and cementless total and unicompartmental knee arthroplasties in patients with low and normal bone mineral density: a 5-year RSA cohort study of 397 cases.Acta orthopaedica · 2026